Agreement between two tympanometers. A methodological study of instrument comparison.

M Gaihede1, F Marker.   

Abstract

When old instruments are replaced by new ones, agreement between them becomes important when comparing measurements. Their precision, too, is important. A common, however inappropriate, statistical approach in these cases is correlation analysis. As an alternative, a simple statistical method is suggested, in which differences between measurements of either instrument are plotted against their means. The distribution of the differences is tested against 0 (one-sample t-test), and for good agreement P > 0.05. This statistical method was used in a group of 32 adults in which measurements taken with an older automatic tympanometer and a new digital one were compared. Static compliance and and gradient showed good agreement, while middle ear pressure showed a significant bias between instruments explained by phase delay. This disagreement was insignificant for clinical purposes, but may be important in scientific studies. The same statistical principle applies determining the precision of each instrument. The new digital instrument showed higher precision than the older one.

Mesh:

Year:  1998        PMID: 9638830     DOI: 10.1080/010503998420351

Source DB:  PubMed          Journal:  Scand Audiol        ISSN: 0105-0397


  2 in total

1.  High frequency tympanometry in neonates with normal otoacoustic emissions: measurements and interpretations.

Authors:  Arvinder Singh Sood; Charanjit Singh Bons; Gursharan Singh Narang
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2012-09-04

2.  Performing tympanometry using smartphones.

Authors:  Justin Chan; Ali Najafi; Mallory Baker; Julie Kinsman; Lisa R Mancl; Susan Norton; Randall Bly; Shyamnath Gollakota
Journal:  Commun Med (Lond)       Date:  2022-06-16
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.