K Håkansson1, R Zubarev, P Håkansson.
Abstract
For reliable confirmation of peptide structures, nozzle-skimmer collisionally induced dissociation was found in many cases to be insufficient, and partial acid hydrolysis was employed as a complementary technique. The utility of combining these fragmentation methods is demonstrated in two examples where the complete sequences of two synthetic peptides (peptide I, MW 2290 and peptide II, MW 1482) were unambiguously determined. In a third example, three different valine deletions in a 2 kDa synthetic peptide were identified and their positions unambiguously established. A home-built electrospray ionization time-of-flight mass spectrometer with orthogonal extraction was used for these analyses. The performance of this instrument with a resolving power of up to 7500, a mass accuracy of < or = 10 ppm, and a detection limit of 1 fmol was shown to be well suited for such studies. As a substitution to conventional external calibration, a more convenient and equally accurate internal 3-point calibration is proposed, based on the low mass ions that are present in almost all peptide spectra.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9633116 DOI: 10.1002/(SICI)1097-0231(19980615)12:11<705::AID-RCM212>3.0.CO;2-3
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419