OBJECTIVE: The objective of this study was to evaluate the accuracy of alveolar crestal bone detection in a comparison of unenhanced and enhanced Sidexis (Siemens Medical Systems, Inc., Bensheim, Germany) digital images with Ektaspeed Plus (Eastman Kodak, Rochester, N.Y.) films by means of receiver operating characteristic analysis. STUDY
DESIGN: More than 100 proximal and furcal areas in the anterior and posterior areas of the mandible and maxilla of each of three tissue-equivalent human skull phantoms were imaged with film, direct digital images, and contrast- and brightness-enhanced digital images. Alveolar crest status was assessed by a consensus panel of three experts who used the Delphi method and evaluated information from all of the imaging modalities concurrently. Five observers assessed all images for the presence or absence of crestal bone loss using a five-point confidence scale. Receiver operating characteristic curves were generated, and calculated areas (Az) were analyzed by means of analysis of variance.
RESULTS: The expert panel determined that of 106 crestal areas, 48 were disease free and 58 exhibited bone loss greater than 2 mm apical to the cemento-enamel junction. Analysis of variance of observer Az scores showed significant differences among readers (p < 0.001) and among readings (p = 0.027), but not among modalities (p = 0.435). Mean Az values for the different modalities were as follows: Sidexis, 0.70; enhanced Sidexis, 0.71; Ektaspeed Plus films, 0.735.
CONCLUSION: The Sidexis digital imaging system was not significantly different from Ektaspeed Plus film for crestal bone evaluation in this in vitro study.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9619682     DOI: 10.1016/s1079-2104(98)90299-2

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  5 in total

1.  Measurement accuracy and perceived quality of imaging systems for the evaluation of periodontal structures.

Authors:  B Güniz Baksi
Journal:  Odontology       Date:  2008-07-27       Impact factor: 2.634

2.  Comparison of diagnostic accuracy of conventional intraoral periapical and direct digital radiographs in detecting interdental bone loss.

Authors:  S R Ashwinirani; Girish Suragimath; H P Jaishankar; Prasad Kulkarni; Shobha C Bijjaragi; Varsha Ajit Sangle
Journal:  J Clin Diagn Res       Date:  2015-02-01

3.  Measurements of simulated periodontal bone defects in inverted digital image and film-based radiograph: an in vitro study.

Authors:  Rafael Scaf de Molon; Juliana Aparecida Najarro Dearo Morais-Camillo; Celso Eduardo Sakakura; Mauricio Goncalves Ferreira; Leonor Castro Monteiro Loffredo; Gulnara Scaf
Journal:  Imaging Sci Dent       Date:  2012-12-23

4.  Comparison of Conventional and Standardized Bone Densitometry around Implants in Periapical Radiographs during a Three Months Period.

Authors:  Asieh Zamani Naser; Shahab Etemadi; Mansour Rismanchian; Mahnaz Sheikhi; Mohammadbagher Tavakoli
Journal:  Dent Res J (Isfahan)       Date:  2011

5.  Comparison between inverted and unprocessed digitized radiographic imaging in periodontal bone loss measurements.

Authors:  Gulnara Scaf; Olívia Morihisa; Leonor de Castro Monteiro Loffredo
Journal:  J Appl Oral Sci       Date:  2007-12       Impact factor: 2.698

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.