Cranial bone flap fixation clamps: compatibility at MR imaging.

F G Shellock1, V J Shellock.   

Abstract

Metallic cranial bone flap fixation clamps were tested for compatibility at 1.5-T magnetic resonance (MR) imaging with standardized techniques. There was no magnetic field attraction, heating was minor, and artifacts involved a relatively small signal void. These data indicate that bone flap fixation clamps will not present risks to patients undergoing MR imaging procedures with MR systems of 1.5 T or less.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9609911     DOI: 10.1148/radiology.207.3.9609911