Literature DB >> 9609345

Computational fluid dynamic analyses to establish design process of centrifugal blood pumps.

Y Miyazoe1, T Sawairi, K Ito, Y Konishi, T Yamane, M Nishida, T Masuzawa, K Takiura, Y Taenaka.   

Abstract

To establish quantitative, efficient design theories for centrifugal blood pumps, computational fluid dynamics (CFD) analyses were compared to the results of flow visualization tests and hemolysis tests, mainly on the Nikkiso centrifugal blood pump. The results turned out to coincide in the velocity vector plots. CFD analysis revealed that the smaller the gap is, the greater the shear stress becomes. This tendency becomes even greater with a radial gap change. Hemolysis study also indicated that the smaller the gap is, the greater the hemolysis. CFD analysis in comparison with hemolysis tests could be a useful index for developing blood pumps in the future.

Entities:  

Mesh:

Year:  1998        PMID: 9609345     DOI: 10.1046/j.1525-1594.1998.06169.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  2 in total

Review 1.  The use of computational fluid dynamics in the development of ventricular assist devices.

Authors:  Katharine H Fraser; M Ertan Taskin; Bartley P Griffith; Zhongjun J Wu
Journal:  Med Eng Phys       Date:  2010-11-13       Impact factor: 2.242

2.  Estimation of changes in dynamic hydraulic force in a magnetically suspended centrifugal blood pump with transient computational fluid dynamics analysis.

Authors:  Toru Masuzawa; Akiko Ohta; Nobuatu Tanaka; Yi Qian; Tomonori Tsukiya
Journal:  J Artif Organs       Date:  2009-09-19       Impact factor: 1.731

  2 in total