Mechanical dysfunction of ventriculoperitoneal shunts caused by calcification of the silicone rubber catheter.

A L Boch1, E Hermelin, C Sainte-Rose, S Sgouros.   

Abstract

OBJECT: The authors studied new and calcified shunt catheters to identify the prevalence of failures caused by aging materials in the shunt. Complications associated with these devices have various origins. Among late complications, fracture or migration of the system is related to the subcutaneous adhesion of the distal tubing in a growing child. A review of a cohort of 64 children who underwent shunt placement in 1980 with barium-impregnated distal catheters showed that 10 of these patients underwent reoperation for complications related to aging of the shunt material. This group represents 15% of the whole series and 30% of those children who were followed for more than 3 years. The true impact of aging of materials on shunt function is probably underestimated.
METHODS: The authors performed physical, chemical, and mechanical analyses of the retrieved aged catheters and also of new catheters, resulting in the following findings: 1) calcifications were observed only on the external surface of the catheter, predominantly in its subcutaneous segment at the level of the neck and anterior chest wall; 2) calcifications contained particles of free silicon and barium sulfate, signifying fragmentation of the polymer; 3) the microstructure of the silicone polymer was modified: microfractures and alteration of the polymeric network were observed; 4) silanol groups were observed on the external surface of the catheter; and 5) the mechanical properties of the silicone rubber were degraded, and the aged catheters were more brittle than the new ones, with ruptures at elongations and fracture energy much lower than that seen in new catheters. Furthermore, in vitro testing with a metastable solution of simulated body fluid demonstrated the critical impact of pH variations in liquid media and surface degradation of the catheters on the precipitation of hydroxylapatite crystals.
CONCLUSIONS: Although most shunt complications can be addressed by better patient management and surgical technique, late complications appear to be partly related to aging of the material. Distal tubing calcifications have been observed in barium-impregnated catheters. The industry recently responded to these observations by introducing plain silicone-coated shunt tubing; further evaluation will show what improvement can be expected.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9609291     DOI: 10.3171/jns.1998.88.6.0975

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  5 in total

Review 1.  Nonfunctional abdominal complications of the distal catheter on the treatment of hydrocephalus: an inflammatory hypothesis? Experience with six cases.

Authors:  Humberto Belem de Aquino; Edmur Franco Carelli; Antonio Guilherme Borges Neto; Carlos Umberto Pereira
Journal:  Childs Nerv Syst       Date:  2006-03-09       Impact factor: 1.475

Review 2.  Biomaterials for the central nervous system.

Authors:  Yinghui Zhong; Ravi V Bellamkonda
Journal:  J R Soc Interface       Date:  2008-09-06       Impact factor: 4.118

Review 3.  Urinary bladder perforation, intra-corporeal knotting, and per-urethral extrusion of ventriculoperitoneal shunt in a single patient: case report and review of literature.

Authors:  Rashim Kataria; Virendra D Sinha; Sanjeev Chopra; Ashok Gupta; Nachiket Vyas
Journal:  Childs Nerv Syst       Date:  2012-12-12       Impact factor: 1.475

4.  Shunt tube calcification as a late complication of ventriculoperitoneal shunting.

Authors:  Abubakr Darrag Salim; Mohammed Awad Elzain; Haddab Ahmed Mohamed; Baha Eldin Mohamed Ibrahim Zayan
Journal:  Asian J Neurosurg       Date:  2015 Jul-Sep

5.  A Retrospective Analysis of Ventriculoperitoneal Shunt Revision Cases of a Single Institute.

Authors:  Man-Kyu Park; Myungsoo Kim; Ki-Su Park; Seong-Hyun Park; Jeong-Hyun Hwang; Sung Kyoo Hwang
Journal:  J Korean Neurosurg Soc       Date:  2015-05-31
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.