Experimental and Monte Carlo dosimetry of the Henschke applicator for high dose-rate 192Ir remote afterloading.

Y Watanabe1, J Roy, P J Harrington, L L Anderson.   

Abstract

We have performed extensive computational and experimental dosimetry of the Henschke applicator with respect to high dose-rate 192Ir brachytherapy using a GAMMAMED remote afterloader. Our goal was to generate clinically useful two- and three-dimensional look-up tables. Dose measurements of the Henschke applicator involved using TLD chips placed in a polystyrene phantom. Monte Carlo simulations were performed using the MCNP code. The computational models included the detailed geometry of 192Ir source, tandem tube, and shielded ovoid. The measured dose rates were corrected for the dependence of TLD sensitivity on the distance of measurement points from the source. Transit dose delivered during source extension to and retraction from a given dwell position was estimated by Monte Carlo simulations, and a correction was applied to the experimental values. For the applicator tandem, the ratio of dose rates obtained by MCNP to those measured by TLD chips ranges from 0.92 to 1.10 with an average of 0.98 and a standard deviation of 0.02. The measured and calculated dose rates at 1 cm on the transverse axis are 1.10 cGy U-1 h-1. For the shielded ovoid, the ratio ranges from 0.88 to 1.16 with an average of 1.00 and a standard deviation of 0.07. Causes of the discrepancy between the Monte Carlo and TLD results were identified. We found that the combined uncertainty of measured dose rates due to these causes is 5.6% for the applicator tandem and 8.4% for the shielded ovoid. Therefore, the results of the Monte Carlo simulation are considered to have been validated by the measurements within the uncertainty involved in the calculation and measurements.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9608485     DOI: 10.1118/1.598255

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Monte Carlo dose calculation of GZP6 (60)Co stepping source based on a matrix shift technique.

Authors:  Mohammad Taghi Bahreyni Toossi; Malihe Abdollahi; Mahdi Ghorbani
Journal:  Rep Pract Oncol Radiother       Date:  2010-12-21

2.  Using matrix summation method for three dimensional dose calculation in brachytherapy.

Authors:  Mahmoud Zibandeh-Gorji; Ali Asghar Mowlavi; Saeed Mohammadi
Journal:  Rep Pract Oncol Radiother       Date:  2012-02-09

3.  Influence of metal of the applicator on the dose distribution during brachytherapy.

Authors:  Chin-Hui Wu; An-Cheng Shiau; Yi-Jen Liao; Hsin-Yu Lin; Yen-Wan Hsueh Liu; Shih-Ming Hsu
Journal:  PLoS One       Date:  2014-08-18       Impact factor: 3.240

4.  HDR Brachytherapy Dose Distribution is Influenced by the Metal Material of the Applicator.

Authors:  Chin-Hui Wu; Yi-Jen Liao; An-Cheng Shiau; Hsin-Yu Lin; Yen-Wan Hsueh Liu; Shih-Ming Hsu
Journal:  Sci Rep       Date:  2015-12-11       Impact factor: 4.379

5.  Dosimetric parameters of three new solid core I-125 brachytherapy sources.

Authors:  Timothy D Solberg; John J DeMarco; Geoffrey Hugo; Robert E Wallace
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.