Literature DB >> 9598261

Monitoring evoked potentials during spinal surgery in one institution.

P H Manninen1.   

Abstract

PURPOSE: To review the experience of one tertiary care institution with somatosensory evoked potential (SSEP) monitoring during spinal surgery in order to assess the ability to monitor and predict neurological outcome effectively.
METHODS: Records of all patients undergoing spinal surgery during 18 mo were retrospectively reviewed. Information from the patient chart included preoperative neurological status, surgical procedure, anaesthetic management, and postoperative neurological outcome. Information regarding the techniques used and interpretation of all SSEP tracings were obtained from evoked potential data sheets completed for each patient. The incidences of clinically important SSEP changes and new postoperative neurological deficits were analysed.
RESULTS: Somatosensory evoked potential monitoring of the lower and upper extremities with non invasive techniques was used in 309 patients undergoing surgery on the cervical (88), thoracic (52), and lumbar spine (169). Thirty seven patients (11%) did not have suitable tracings for interpretation and 17 (5.5%) had baseline tracings described as poor. An intraoperative SSEP change occurred in 16 patients (6%) with SSEP and seven (2.6%) had a new neurological deficit postoperatively. Three persistent deficits were predicted by permanent SSEP change, and one transient deficit by a transient SSEP change. False positive results occurred in 12 patients (4.4%) and false negative results occurred in three (1.1%), with a sensitivity of 57% and a specificity of 95%. The incidence of SSEP changes was greater in the thoracic (18%) than in the cervical (1.2%) or lumbar (5.4%) groups (P < 0.05).
CONCLUSION: Effective SSEP monitoring was possible despite the many factors which may have interfered with monitoring. More improvements in the techniques and conditions of monitoring are needed to decrease the incidence of false positive and negative results.

Entities:  

Mesh:

Year:  1998        PMID: 9598261     DOI: 10.1007/BF03012582

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  6 in total

Review 1.  False negative findings in intraoperative SEP monitoring: analysis of 658 consecutive neurosurgical cases and review of published reports.

Authors:  H Wiedemayer; I E Sandalcioglu; W Armbruster; J Regel; H Schaefer; D Stolke
Journal:  J Neurol Neurosurg Psychiatry       Date:  2004-02       Impact factor: 10.154

2.  Representation of somatosensory evoked potentials using discrete wavelet transform.

Authors:  Ulrich Hoppe; Kai Schnabel; Stephan Weiss; Ingrid Rundshagen
Journal:  J Clin Monit Comput       Date:  2002 Apr-May       Impact factor: 2.502

3.  Predictive factors for neurological deterioration after surgical decompression for thoracic ossified yellow ligament.

Authors:  Chris Yuk Kwan Tang; Jason Pui Yin Cheung; Dino Samartzis; Ka Hei Leung; Yat Wa Wong; Keith Dip Kei Luk; Kenneth Man Chee Cheung
Journal:  Eur Spine J       Date:  2017-04-03       Impact factor: 3.134

4.  Sciatic nerve injury associated with acetabular fractures.

Authors:  Paul S Issack; David L Helfet
Journal:  HSS J       Date:  2008-12-17

5.  Transcranial Motor Evoked Potentials during Spinal Deformity Corrections-Safety, Efficacy, Limitations, and the Role of a Checklist.

Authors:  Shankar Acharya; Nagendra Palukuri; Pravin Gupta; Manish Kohli
Journal:  Front Surg       Date:  2017-02-13

6.  A review of intraoperative monitoring for spinal surgery.

Authors:  Mark M Stecker
Journal:  Surg Neurol Int       Date:  2012-07-17
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.