A survey of the efficiency of visible light curing units.

F E Martin1.   

Abstract

OBJECTIVES: The purpose of this study was to survey the efficiency of visible light curing units in dental practices across Australia.
METHODS: Survey forms were distributed to representatives of 3M Health Care to complete when visiting dentists in their working areas. The information collected included the type and age of the unit, curing times used, history of maintenance, replacement of components, and the light intensity reading.
RESULTS: Of the 214 light curing units surveyed, approximately 27% recorded a light intensity of 200 mW cm-2 or less, a level regarded as inadequate to cure a 2-mm thick increment of composite resin. An additional 26% registered an output of between 201 and 399 mW m-2. This level would be considered acceptable with additional curing time; however, 44% of practitioners were curing for 20 s or less. A negative correlation was found between the age of the unit and the intensity recorded. Nearly 50% of respondents had never checked the light output of their unit.
CONCLUSIONS: The results indicate that just over one-half of the light curing units surveyed were not functioning satisfactorily. An obvious reduction in intensity was noted with the older units. There is a substantial lack of awareness among dentists of the need for maintenance and regular checking of the light intensity of these units.

Mesh:

Substances:

Year:  1998        PMID: 9594476     DOI: 10.1016/s0300-5712(97)00004-3

Source DB:  PubMed          Journal:  J Dent        ISSN: 0300-5712            Impact factor:   4.379


  12 in total

1.  Influence of physical assessment of different light-curing units on irradiance and composite microhardness top/bottom ratio.

Authors:  Susana Morimoto; Renata Aló Maluza Zanini; Josete Barbosa Cruz Meira; Carlos Martins Agra; Fernanda Calabró Calheiros; Denis Yudi Nagase
Journal:  Odontology       Date:  2016-01-13       Impact factor: 2.634

2.  Evaluation of the curing depth of two translucent composite materials using a halogen and two LED curing units.

Authors:  Olga Polydorou; Alexandros Manolakis; Elmar Hellwig; Petra Hahn
Journal:  Clin Oral Investig       Date:  2007-10-25       Impact factor: 3.573

3.  A survey of power density of light-curing units used in private dental offices in Changchun City, China.

Authors:  Xinqing Hao; Meng Luo; Jian Wu; Song Zhu
Journal:  Lasers Med Sci       Date:  2013-05-23       Impact factor: 3.161

4.  Characterization of heat emission of light-curing units.

Authors:  Mohammed A Wahbi; F A Aalam; F I Fatiny; S A Radwan; I Y Eshan; K H Al-Samadani
Journal:  Saudi Dent J       Date:  2012-03-09

5.  Light intensity decay in quartz-tungsten-halogen polymerization units.

Authors:  Khalid H Al-Samadani; Al-Dharrab Ayman; A Wahbi Mohammed; I Algizani Loay
Journal:  J Int Oral Health       Date:  2013-02-26

6.  The Effect of Different Light Curing Units and Tip Distances on Translucency Parameters of Bulk fill Materials.

Authors:  Dana Al-Senan; Hend Al-Nahedh
Journal:  Saudi Dent J       Date:  2022-04-12

7.  Effects of light curing method and exposure time on mechanical properties of resin based dental materials.

Authors:  A Riza Alpöz; Fahinur Ertugrul; Dilsah Cogulu; Asli Topaloglu Ak; Metin Tanoglu; Elçin Kaya
Journal:  Eur J Dent       Date:  2008-01

8.  Knowledge and Attitude of Dental Clinicians towards Light-Curing Units: A Cross-Sectional Study.

Authors:  Dana Al-Senan; Fatin Ageel; Amani Aldosari; Haifa Maktabi
Journal:  Int J Dent       Date:  2021-06-14

9.  The effect of different light-curing units on tensile strength and microhardness of a composite resin.

Authors:  Eduardo Batista Franco; Patrícia Aleixo dos Santos; Rafael Francisco Lia Mondelli
Journal:  J Appl Oral Sci       Date:  2007-12       Impact factor: 2.698

10.  Depth of cure of dental composites submitted to different light-curing modes.

Authors:  Raphael Vieira Monte Alto; José Guilherme Antunes Guimarães; Laiza Tatiana Poskus; Eduardo Moreira da Silva
Journal:  J Appl Oral Sci       Date:  2006-04       Impact factor: 2.698

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.