Serum protein electrophoresis and immunofixation by a semiautomated electrophoresis system.

X Bossuyt1, A Bogaerts, G Schiettekatte, N Blanckaert.   

Abstract

Semiautomated agarose electrophoresis and immunofixation performed with Hydrasys-Hyrys (Sebia) were compared with conventional, manually performed methods, including cellulose acetate electrophoresis, immunoelectrophoresis, and immunofixation. Reference intervals for agarose electrophoresis with Hydrasys-Hyrys were determined. Within-run imprecision (CV) for fraction quantitation with the semiautomated system was between 1% (albumin) and 4.5% (beta-globulin). Total imprecision (CV) was between 2.7% (albumin) and 7.3% (beta-globulin). Semiautomated agarose electrophoresis showed linear correlation with cellulose acetate electrophoresis. Thirty-four specimens with monoclonal components were analyzed by manual immunoelectrophoresis and immunofixation and by Hydrasys. In one case, a light-chain disease was missed with Hydrasys when the sample was diluted 1:3 (the routine dilution) but not when the sample was assayed undiluted. In another case, the Hydrasys system revealed a small IgGA monoclonal component in addition to the IgA monoclonal component detected by the manual methods. In the other cases, no differences between the manual methods and the semiautomated method were seen with respect to paraprotein identification.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9590366

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  3 in total

1.  Blood monitoring of circulating tumor plasma cells by next generation flow in multiple myeloma after therapy.

Authors:  Luzalba Sanoja-Flores; Juan Flores-Montero; Noemi Puig; Teresa Contreras-Sanfeliciano; Roberia Pontes; Alba Corral-Mateos; Omar García-Sánchez; María Díez-Campelo; Roberto José Pessoa de Magalhães; Luis García-Martín; José María Alonso-Alonso; Aranzazú García-Mateo; Carlos Aguilar-Franco; Jorge Labrador; Abelardo Barez-García; Angelo Maiolino; Bruno Paiva; Jesús San Miguel; Elaine Sobral da Costa; Marcos González; María Victoria Mateos; Brian Durie; Jacques J M van Dongen; Alberto Orfao
Journal:  Blood       Date:  2019-12-12       Impact factor: 22.113

2.  Accurate identification of paraprotein antigen targets by epitope reconstruction.

Authors:  Seshi R Sompuram; Gerassimos Bastas; Kodela Vani; Steven A Bogen
Journal:  Blood       Date:  2007-09-18       Impact factor: 22.113

3.  Immunoglobulin Heavy Chain Gene Rearrangements in Patients with Gaucher Disease.

Authors:  Predrag Rodić; Milan Lakočević; Sonja Pavlović; Teodora Karan Đurašević; Tatjana Kostić; Nada Suvajdžić Vuković; Zorica Šumarac; Milan Petakov; Dragana Janić
Journal:  J Med Biochem       Date:  2018-07-01       Impact factor: 3.402

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.