OBJECTIVE: To ascertain the microbial load and type of organisms on used surgical instruments following standard cleaning, which consisted of the use of a washer sterilizer followed by sonic cleaning. DESIGN: In this prospective experimental study, used surgical instruments were immersed in Peptamin Tween broth, the broth agitated, and then filtered through a 0.45 microm filter. Quantitative cultures were performed, and all microbes were identified by using standard techniques. SETTING: This study was conducted at a 660-bed university hospital. RESULTS: The microbial load remaining on used surgical instruments after cleaning was as follows: 36 (72%) instruments 0 to 10 colony-forming units (CFU), 7 (14%) instruments 11 to 100 CFU, and 7 (14%) instruments > 100 CFU. Organisms contaminating the instruments included coagulase-negative staphylococcus (56%) followed by Bacillus (22%) and diphtheroids (14%). No other microbes were isolated from more than 4% of the instruments. CONCLUSION: Most used nonlumen surgical instruments contain less than 100 CFU of relatively nonpathogenic microorganisms after cleaning. This suggests that new low-temperature sterilization technologies are likely to be highly effective in preventing cross-transmission of infection via nonlumen medical instruments.
OBJECTIVE: To ascertain the microbial load and type of organisms on used surgical instruments following standard cleaning, which consisted of the use of a washer sterilizer followed by sonic cleaning. DESIGN: In this prospective experimental study, used surgical instruments were immersed in Peptamin Tween broth, the broth agitated, and then filtered through a 0.45 microm filter. Quantitative cultures were performed, and all microbes were identified by using standard techniques. SETTING: This study was conducted at a 660-bed university hospital. RESULTS: The microbial load remaining on used surgical instruments after cleaning was as follows: 36 (72%) instruments 0 to 10 colony-forming units (CFU), 7 (14%) instruments 11 to 100 CFU, and 7 (14%) instruments > 100 CFU. Organisms contaminating the instruments included coagulase-negative staphylococcus (56%) followed by Bacillus (22%) and diphtheroids (14%). No other microbes were isolated from more than 4% of the instruments. CONCLUSION: Most used nonlumen surgical instruments contain less than 100 CFU of relatively nonpathogenic microorganisms after cleaning. This suggests that new low-temperature sterilization technologies are likely to be highly effective in preventing cross-transmission of infection via nonlumen medical instruments.
Authors: Curtis J Donskey; Marian Yowler; Yngve Falck-Ytter; Sirisha Kundrapu; Robert A Salata; William A Rutala Journal: Antimicrob Resist Infect Control Date: 2014-01-21 Impact factor: 4.887