The self-inflating bulb to detect esophageal intubation during emergency airway management.

C L Kasper1, S Deem.   

Abstract

BACKGROUND: The negative-pressure test using a self-inflating bulb (SIB) during emergency intubation was studied to determine its reliability and predictive value in this setting.
METHODS: The endotracheal tube (ETT) position was tested in 300 consecutive patients undergoing in-hospital emergency endotracheal intubation. Immediately after intubation and before ETT cuff inflation, the following protocol was strictly followed: (1) an SIB was compressed, connected to the ETT, and released. A 10-s period was allowed for the bulb to inflate. (2) The ETT cuff was inflated, and the ETT position was confirmed using colorimetric or infrared carbon dioxide detection, or both, combined with clinical evaluation.
RESULTS: There were 19 esophageal intubations (6% incidence). The SIB correctly identified all patients with esophageal intubation (sensitivity, 100%) and correctly identified all but three ETTs placed in the trachea (specificity, 99%). The three tracheally placed tubes that were misidentified by the bulb syringe occurred during one case each of chronic obstructive pulmonary disease, copious secretions, and obesity; of note were three tracheally placed tubes that were misidentified by the carbon dioxide analyzers during cardiopulmonary resuscitation.
CONCLUSIONS: The SIB proved to be a sensitive and specific test for esophageal intubation in the emergency setting when used according to the protocol described, and it is complementary to carbon dioxide detection. The predictive value of the bulb syringe appears to be improved when a prolonged period for reinflation is allowed. It holds particular promise because of its low cost and portability.

Entities:  

Mesh:

Year:  1998        PMID: 9579497     DOI: 10.1097/00000542-199804000-00008

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  2 in total

1.  A novel airway device with tactile sensing capabilities for verifying correct endotracheal tube placement.

Authors:  Pauwel Goethals; Harshu Chaobal; Dominiek Reynaerts; David Schaner
Journal:  J Clin Monit Comput       Date:  2013-09-25       Impact factor: 2.502

2.  Canadian Airway Focus Group updated consensus-based recommendations for management of the difficult airway: part 1. Difficult airway management encountered in an unconscious patient.

Authors:  J Adam Law; Laura V Duggan; Mathieu Asselin; Paul Baker; Edward Crosby; Andrew Downey; Orlando R Hung; Philip M Jones; François Lemay; Rudiger Noppens; Matteo Parotto; Roanne Preston; Nick Sowers; Kathryn Sparrow; Timothy P Turkstra; David T Wong; George Kovacs
Journal:  Can J Anaesth       Date:  2021-06-18       Impact factor: 5.063

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.