| Literature DB >> 9570103 |
V Gomez Escobar1, F Vera Tomé, J C Lozano, A Martín Sánchez.
Abstract
An extractive procedure for uranium determination using liquid scintillation counting with the URAEX cocktail is described. Interference from radon and a strong influence of nitrate ion were detected in this procedure. Interference from radium, thorium and polonium emissions were very low when optimal operating conditions were reached. Quenching effects were considered and the minimum detectable activity was evaluated for different sample volumes. Isotopic analysis of samples can be performed using the proposed method. Comparisons with the results obtained with the general procedure used in alpha spectrometry with passivated implanted planar silicon detectors showed good agreement. The proposed procedure is thus suitable for uranium determination in water samples and can be considered as an alternative to the laborious conventional chemical preparations needed for alpha spectrometry methods using semiconductor detectors.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9570103 DOI: 10.1016/s0969-8043(97)10001-x
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513