Literature DB >> 9561766

Solid-phase microextraction of polychlorinated biphenyls.

Y Yang1, D J Miller, S B Hawthorne.   

Abstract

The extraction and analysis of 21 polychlorinated biphenyls (PCBs) ranging from di- to decachlorobiphenyls in ocean, wetland and leachate water samples were achieved using solid-phase microextraction (SPME) with a 100-micron poly(di-methylsiloxane) (PDMS) fiber and gas chromatography-electron-capture detection (GC-ECD). Severe carryover between samples (e.g., 20%) occurs on both stir bars and the SPME fibers demonstrating that it is important to use a new stir bar for each sample, as well as to perform SPME-GC blanks between samples to avoid quantitative errors. The equilibrium partitioning coefficients of individual PCB congeners between PDMS and water were found to be surprisingly different compared to their octanol-water partitioning coefficient (Kow), demonstrating that Kow cannot be used to estimate the partitioning behavior of PCBs in the SPME process. Using a 15-min SPME extraction, SPME analysis with GC-ECD was linear (r2 > or = 0.97) from approximately 5 pg/ml to the solubility limit of each congener. Concentrations in water samples obtained by 15-min SPME extractions compared favorably with those obtained by toluene extractions, demonstrating that SPME combined with GC is a useful technique for the rapid determination of PCBs in water samples.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9561766     DOI: 10.1016/s0021-9673(97)01115-1

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  3 in total

1.  Simultaneous analysis of organochlorine pesticides and polychlorinated biphenyls in air samples by using accelerated solvent extraction (ASE) and solid-phase micro-extraction (SPME) coupled to gas chromatography dual electron capture detection.

Authors:  Haifaa Mokbel; Enaam Jamal Al Dine; Ahmad Elmoll; Céline Liaud; Maurice Millet
Journal:  Environ Sci Pollut Res Int       Date:  2016-01-16       Impact factor: 4.223

2.  Extraction of ultra traces of polychlorinated biphenyls in aqueous samples using suspended liquid-phase microextraction and gas chromatography-electron capture detection.

Authors:  Jalal Hassan; Mojtaba Shamsipur
Journal:  Environ Monit Assess       Date:  2012-08-16       Impact factor: 2.513

3.  Quantification of PAHs and chlorinated compounds by novel solid-phase microextraction based on the arrays of tin oxide nanorods.

Authors:  Reza Alizadeh; Nahid Mashkouri Najafi
Journal:  Environ Monit Assess       Date:  2013-01-30       Impact factor: 2.513