Maintaining fixed band spacing when changing column dimensions in gradient elution.

J W Dolan1, L R Snyder.   

Abstract

In gradient elution separations, it may be required to change either column length (to increase resolution or shorten run time) or column diameter (for an increase in sensitivity or for preparative separations). In either of these changes of column dimensions, it is usually desired to maintain the same relative band spacing (selectivity), so as to increase resolution in proportion to (column plate number)1/2 when increasing column length, or to maintain constant resolution when changing column diameter. A general rule for avoiding changes in band spacing in these situations is to maintain the quantity [(gradient time) x (flow-rate)/(column volume)] constant, while holding the initial and final gradient mobile phase compositions (%B) fixed. This rule is only valid as long as the equipment hold-up volume (dwell volume) is negligible, or if all sample components are strongly retained at the start of the gradient. When neither of the latter conditions apply, then significant changes in band spacing may result when changing column size. Rules are presented for recognizing this potential problem for a given sample/HPLC-equipment combination, and adjustments in separation conditions that can avoid this problem are discussed. Changes in band spacing as a result of change in column size are of special concern when developing procedures for preparative chromatography under gradient conditions.

Mesh:

Substances:

Year:  1998        PMID: 9550100     DOI: 10.1016/s0021-9673(97)01045-5

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  4 in total

Review 1.  Fast, comprehensive two-dimensional liquid chromatography.

Authors:  Dwight R Stoll; Xiaoping Li; Xiaoli Wang; Peter W Carr; Sarah E G Porter; Sarah C Rutan
Journal:  J Chromatogr A       Date:  2007-08-30       Impact factor: 4.759

2.  High-speed gradient elution reversed-phase liquid chromatography of bases in buffered eluents. Part I. Retention repeatability and column re-equilibration.

Authors:  Adam P Schellinger; Dwight R Stoll; Peter W Carr
Journal:  J Chromatogr A       Date:  2008-01-31       Impact factor: 4.759

3.  Instrument parameters controlling retention precision in gradient elution reversed-phase liquid.

Authors:  Ayse Beyaza; Wenzhe Fana; Peter W Carr; Adam P Schellinger
Journal:  J Chromatogr A       Date:  2014-12-05       Impact factor: 4.759

4.  "Measure Your Gradient": a new way to measure gradients in high performance liquid chromatography by mass spectrometric or absorbance detection.

Authors:  Megan H Magee; Joseph C Manulik; Brian B Barnes; Daniel Abate-Pella; Joshua T Hewitt; Paul G Boswell
Journal:  J Chromatogr A       Date:  2014-10-08       Impact factor: 4.759

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.