Endoleak after endovascular graft repair of experimental aortic aneurysms: does coil embolization with angiographic "seal" lower intraaneurysmal pressure?

B Marty1, L A Sanchez, T Ohki, R A Wain, P L Faries, J Cynamon, M L Marin, F J Veith.   

Abstract

PURPOSE: To investigate the relation between endoleaks and intraaneurysmal pressure (IAP) and the effect of coil embolization in the management of endoleaks.
METHODS: The infrarenal aorta of a dog (n = 15) was replaced by a polytetrafluoroethylene aneurysm containing a pressure transducer. Group I (n = 4) had untreated aneurysms. Group II (n = 4) had endovascularly excluded aneurysms without an endoleak. Group III (n = 7) had aneurysms excluded by means of grafts with a defect that represented the source of an endoleak. After 4 weeks of follow-up study, the endoleaks in group III dogs were subjected to coil embolization. Systolic IAP was measured daily and expressed as a ratio of systolic blood pressure obtained from a forelimb cuff. Arteriography, duplex ultrasonography, and spiral contrast computed tomography were performed to evaluate endoleaks.
RESULTS: In group I, the LAP remained close to systolic blood pressure (ratio of 0.96 +/- 0.06), whereas in group II the IAP ratio showed a decline to 0.34 +/- 0.16 (p = 0.0009 group I versus II). After an initial decrease, the IAP ratio in group III stabilized at 0.75 +/- 0.18 (p = 0.003, group II versus III). Aneurysms with an endoleak remained pulsatile with a pulse pressure of 30 +/- 16 mm Hg, which was less than that of untreated aneurysms (62 +/- 15 mm Hg; p < 0.0001 group I versus III). Arteriography and computed tomography revealed "sealing" of endoleaks after coil embolization, but IAP ratio did not decrease (0.76 +/- 0.14) after coil embolization.
CONCLUSIONS: Incomplete endovascular aneurysm exclusion caused by an endoleak fails to reduce IAP ratio and may subject the aneurysm to a continued risk for rupture. Although coil embolization resulted in angiographic and computed tomographic sealing, it failed to reduce IAP ratio.

Entities:  

Mesh:

Year:  1998        PMID: 9546230     DOI: 10.1016/s0741-5214(98)70320-9

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  3 in total

1.  Increasing incidence of midterm and long-term complications after endovascular graft repair of abdominal aortic aneurysms: a note of caution based on a 9-year experience.

Authors:  T Ohki; F J Veith; P Shaw; E Lipsitz; W D Suggs; R A Wain; M Bade; M Mehta; N Cayne; J Cynamon; J Valldares; J McKay
Journal:  Ann Surg       Date:  2001-09       Impact factor: 12.969

2.  Endovascular stent-grafting of anastomotic pseudoaneurysms following thoracic aortic surgery.

Authors:  Toshiro Ito; Yoshihiko Kurimoto; Nobuyoshi Kawaharada; Tetsuya Koyanagi; Hitoki Hashiguchi; Akitatsu Yamashita; Yasuko Miyaki; Akihiko Yamauchi; Masanori Nakamura; Tetsuya Higami
Journal:  Gen Thorac Cardiovasc Surg       Date:  2009-10-16

Review 3.  Endotension: twenty years of a controversial term.

Authors:  Álvaro Torres-Blanco; Manuel Miralles-Hernández
Journal:  CVIR Endovasc       Date:  2021-06-05
  3 in total