Total radiated power, infrared output, and heat generation by cold light sources at the distal end of endoscopes and fiber optic bundle of light cables.

C Hensman1, G B Hanna, T Drew, H Moseley, A Cuschieri.   

Abstract

BACKGROUND: Skin burns and ignition of drapes have been reported with the use of cold light sources. The aim of the study was to document the temperature generated by cold light sources and to correlate this with the total radiated power and infrared output.
METHODS: The temperature, total radiated power, and infrared output were measured as a function of time at the end of the endoscope (which is inserted into the operative field) and the end of the fiber optic bundle of the light cable (which connects the cable to the light port of the endoscope) using halogen and xenon light sources.
RESULTS: The highest temperature recorded at the end of the endoscope was 95 degrees C. The temperature measured at the optical fiber location of the endoscope was higher than at its lens surface (p < 0.0001). At the end of the fiber optic bundle of light cables, the temperature reached 225 degrees C within 15 s. The temperature recorded at the optical fiber location of all endoscopes and light cables studied rose significantly over a period of 10 min to reach its maximum (p <0.0001) and then leveled off for the duration of the study (30 min). The infrared output accounted only for 10% of the total radiated power.
CONCLUSIONS: High temperatures are reached by 10 min at the end of fiber optic bundle of light cables and endoscopes with both halogen and xenon light sources. This heat generation is largely due to the radiated power in the visible light spectrum.

Entities:  

Mesh:

Year:  1998        PMID: 9543524     DOI: 10.1007/s004649900665

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  7 in total

Review 1.  Comprehensive review on endonasal endoscopic sinus surgery.

Authors:  Rainer K Weber; Werner Hosemann
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2015-12-22

2.  Solid-state semiconductors are better alternatives to arc-lamps for efficient and uniform illumination in minimal access surgery.

Authors:  Alex C H Lee; Daniel S Elson; Mark A Neil; Sunil Kumar; Bingo W Ling; Fernando Bello; George B Hanna
Journal:  Surg Endosc       Date:  2008-03-25       Impact factor: 4.584

3.  Peritoneal transforming growth factor beta-1 expression during laparoscopic surgery: a clinical trial.

Authors:  Walter J A Brokelman; Lena Holmdahl; Maria Bergström; Peter Falk; Jean H G Klinkenbijl; Jean H G Klinkonbijl; Michel M P J Reijnen; Michael M P J Reijnen
Journal:  Surg Endosc       Date:  2007-03-01       Impact factor: 4.584

4.  Thermal Safety of Endoscopic Usage in Robot-Assisted Middle Ear Surgery: An Experimental Study.

Authors:  Jinxi Pan; Haoyue Tan; Jun Shi; Zhaoyan Wang; Olivier Sterkers; Huan Jia; Hao Wu
Journal:  Front Surg       Date:  2021-05-14

Review 5.  Current State of Surgical Lighting.

Authors:  Jahnavi Curlin; Charles K Herman
Journal:  Surg J (N Y)       Date:  2020-06-19

Review 6.  Vocal fold injury models in rats: a literature review on techniques and methodology.

Authors:  Peter Laszlo Ujvary; Cristina Maria Blebea; Alma Aurelia Maniu; Sever Pop; Orsolya Sarpataki; Marcel Cosgarea
Journal:  J Med Life       Date:  2022-03

7.  Low-Dose, Long-Wave UV Light Does Not Affect Gene Expression of Human Mesenchymal Stem Cells.

Authors:  Darice Y Wong; Thanmayi Ranganath; Andrea M Kasko
Journal:  PLoS One       Date:  2015-09-29       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.