Asymmetric sampling along k(slice-select) in two-dimensional multislice MRI.

J G Pipe1.   

Abstract

Reduction of the slice-select refocusing gradient in two-dimensional multislice imaging results in asymmetry of the k-space representation of collected data along the slice-select direction. Standard methods of partial Fourier reconstruction developed for other methods of asymmetric k-space sampling can be used to reconstruct these data with final through-plane resolution smaller than the collected slice thickness. This method can be used for reducing scan time in the same manner as asymmetric sampling in the phase-encoded direction. In addition, the reduced refocusing gradient reduces minimum TE and motion artifacts in the same manner as for asymmetric sampling in the frequency-encoded direction (fractional echoes). Results using a resolution phantom and a flow phantom illustrate these concepts.

Mesh:

Year:  1998        PMID: 9543425     DOI: 10.1002/mrm.1910390416

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668