Comparison of in-vivo body composition using two Lunar dual-energy X-ray absorptiometers.

B Oldroyd1, J G Truscott, G Woodrow, R Milner, S P Stewart, A H Smith, C F Westmacott, M A Smith.   

Abstract

OBJECTIVE: To compare in-vivo composition analysis between two dual energy X-ray absorptiometers, a DPX and a DPX/L, from the same manufacturer (LUNAR), pre(Study A) and post(Study B) hardware changes on both absorptiometers.
DESIGN: Comparison of (1) quality assurance (QA) data: air-counts low (38 keV), air-counts high (70 keV), air-counts ratio, percent spillover, R-delrin; and (2) total body compartments: total body tissue (TBTISS), total body fat (TBF), percent total body fat (%TBF), total body lean (TBLEAN), total body bone mineral content (TBBMC) and total body bone mineral density (TBBMD), between the two absorptiometers.
SETTING: Centre for Bone and Body Composition Research, University of Leeds.
SUBJECTS: Study A, 14 normal subjects and Study B, a different cohort of 19 normal subjects, were scanned on both machines on the same day.
RESULTS: In Study A, large significant differences were observed in the QA parameters between the two machines. The DPX, air-counts low and air-counts high, being 25% and 22% lower than the DPX/L. The Bland Altman method of analysis indicated that the DPX was significantly higher for TBTISS (0.3 kg), %TBF (2%) and TBF ( 1.4 kg) and correspondingly lower for TBLEAN (-1.0 kg). No significant difference was observed in TBBMC. After the hardware changes (Study B) a marked reduction in the differences in QA air-counts was observed. The DPX air-counts low was now 1% higher and air-counts high 8% lower than the DPX/L. The DPX had now only small significant negative differences for %TBF (-0.6%) and TBF (-0.4 kg) and a small significant positive difference for TBLEAN (0.4 kg), compared to the DPX/L. TBBMC difference although slightly increased, was still non-significant.
CONCLUSIONS: The closer agreement observed in the QA parameters after the hardware changes was associated with a reduction in the mean differences, 95%CI of the mean differences and limits of agreement of the comparison of body composition analysis from the Lunar machines using the Bland-Altman method. The study indicates that the QA limits set for bone mineral analysis may require more stringent limits for body composition.

Entities:  

Mesh:

Year:  1998        PMID: 9537303     DOI: 10.1038/sj.ejcn.1600537

Source DB:  PubMed          Journal:  Eur J Clin Nutr        ISSN: 0954-3007            Impact factor:   4.016


  2 in total

1.  iDXA, Prodigy, and DPXL dual-energy X-ray absorptiometry whole-body scans: a cross-calibration study.

Authors:  Holly Hull; Qing He; John Thornton; Fahad Javed; Lynn Allen; Jack Wang; Richard N Pierson; Dympna Gallagher
Journal:  J Clin Densitom       Date:  2008-11-22       Impact factor: 2.617

2.  Body Fat Mass Assessment: A Comparison between an Ultrasound-Based Device and a Discovery A Model of DXA.

Authors:  Jean-Claude Pineau; Loïc Lalys; Massimo Pellegrini; Nino Carlo Battistini
Journal:  ISRN Obes       Date:  2013-02-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.