A rate-controlled indentor for in vivo analysis of residual limb tissues.

A P Pathak1, M B Silver-Thorn, C A Thierfelder, T E Prieto.   

Abstract

A tissue tester was designed to enable rate-controlled indentation of the bulk soft tissues of lower extremity residual limbs. The tissue tester employs a digital linear actuator that implements rate-controlled indentation, and a load cell that measures the reaction force resulting from tissue indentation. Viscoelastic phenomena such as preconditioning, hysteresis and force relaxation can be assessed, and the effect of varying indentation rates on soft tissue stiffness can be investigated. The device accommodates indentor excursions up to 30 mm, indentation at rates of 0 to 10 mm/s, reaction forces up to 44 N, and multiple loading/unloading cycles. The tissue tester is controlled via a notebook personal computer with a PCMCIA data acquisition card. The tissue testing trials are automated and the entire test system is portable and amenable for use in a clinical or research environment. System output consists of force-displacement curves from cyclic loading, and force-time curves following ramped-step indentation. The mean indentor positioning error was 0.071 (+/-0.75)% of the desired displacement. This error varied as a function of indentation and was approximately independent of the indentation rate. Indentation rates were accurate to within 0.94(+/-0.68)% of the desired value and also varied with indentation. Indentation of a viscoelastic foam yielded force-displacement curves that were consistent with that obtained from an Instron universal testing machine.

Mesh:

Year:  1998        PMID: 9535519     DOI: 10.1109/86.662616

Source DB:  PubMed          Journal:  IEEE Trans Rehabil Eng        ISSN: 1063-6528


  2 in total

1.  Assessment of neck tissue fibrosis using an ultrasound palpation system: a feasibility study.

Authors:  Y P Zheng; S F Leung; A F Mak
Journal:  Med Biol Eng Comput       Date:  2000-09       Impact factor: 2.602

2.  Combined compression and elongation experiments and non-linear modelling of liver tissue for surgical simulation.

Authors:  C Chui; E Kobayashi; X Chen; T Hisada; I Sakuma
Journal:  Med Biol Eng Comput       Date:  2004-11       Impact factor: 2.602

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.