Literature DB >> 9535367

Eradicating acute hemolysis following transcatheter closure of ductus arteriosus by immediate deployment of a second device.

L H Wang1, J K Wang, C E Mullins.   

Abstract

Two patients who underwent transcatheter closure of patent ductus arteriosus, one with a Rashkind umbrella device and the other with a coil, suffered from acute hemolysis following the procedure. Hemolysis ceased after deployment of second device(s) within 48 hr without needing to retrieve the first devices in either patient. We conclude that immediate deployment of a second device(s) is an alternative to surgery when acute hemolysis occurs following transcatheter closure of ductus.

Entities:  

Mesh:

Year:  1998        PMID: 9535367     DOI: 10.1002/(sici)1097-0304(199803)43:3<295::aid-ccd11>3.0.co;2-l

Source DB:  PubMed          Journal:  Cathet Cardiovasc Diagn        ISSN: 0098-6569


  1 in total

1.  Transcatheter occlusion of the patent ductus arteriosus: a comparison of two devices.

Authors:  M K El Mallah; A J Sands; F A Casey; B G Craig; H C Mulholland
Journal:  Ir J Med Sci       Date:  2002 Jul-Sep       Impact factor: 1.568

  1 in total