Literature DB >> 9531157

The "Wedge" resection device for electrosurgical transurethral prostatectomy.

A P Perlmutter1, D A Schulsinger.   

Abstract

The "Wedge" (Microvasive, Natick MA) is a new electroresection device for transurethral prostatectomy (TURP) using the standard resectoscope. The design, which is broader than the standard loop and thickens from front to back, results in better hemostasis when used at 275 to 300 W because of its ability to cut and coagulate tissue simultaneously. In the canine model, histologic examination demonstrated a 2-mm zone of coagulation around the chips and in the resection bed; this response was not observed in the specimens resected by the standard tungsten loop. No adjacent tissue damage was found with either the Wedge or the loop, and the temperatures recorded at the capsule rose only 4 degrees C regardless of the device used. In the 65 patients treated, the average hematocrit drop on postoperative Day 1 was 3.0%, and serum sodium was unchanged. One year postoperatively, the peak flow rate had increased by 101%, and the AUA Symptom Score was 6.1. The only surgical complication was urethral strictures (3%) necessitating incision. Most striking was the increased case of resection attributable to improved intraoperative vision. The data suggest that Wedge TURP is as safe and efficacious as standard loop TURP. The surgical field is markedly improved and clear because of intraoperative hemostasis. A TURP can be performed with a view toward minimizing patient morbidity and increasing surgical ease.

Entities:  

Mesh:

Year:  1998        PMID: 9531157     DOI: 10.1089/end.1998.12.75

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  7 in total

Review 1.  Transurethral vaporization of the prostate: current techniques.

Authors:  M A Cabelin; A E Te; S A Kaplan
Journal:  Curr Urol Rep       Date:  2000-08       Impact factor: 3.092

2.  Comparison of transurethral prostate resection and plasmakinetic prostate resection according to cautery artefacts in tissue specimens.

Authors:  K Turgay Akgül; Ali Ayyildiz; Bariş Nuhoğlu; Muzaffer Caydere; Hüseyin Ustün; Cankon Germiyanoğlu
Journal:  Int Urol Nephrol       Date:  2007-03-13       Impact factor: 2.370

3.  Comparison of TURP, TUVRP, and HoLEP.

Authors:  Narmada P Gupta; Ajay Anand
Journal:  Curr Urol Rep       Date:  2009-07       Impact factor: 3.092

4.  Changes of serum prostate-specific antigen following high energy thick loop prostatectomy.

Authors:  R F Talic; A M El-Tiraifi; S Altaf; T B Hargreave
Journal:  Int Urol Nephrol       Date:  2000       Impact factor: 2.370

5.  The learning curve for holmium laser enucleation of the prostate: a single-center experience.

Authors:  Jungbum Bae; Seung-June Oh; Jae-Seung Paick
Journal:  Korean J Urol       Date:  2010-10-21

6.  Plasmakinetic vaporization versus transurethral resection of the prostate: Six-year results.

Authors:  Orhan Koca; Muzaffer Oğuz Keleş; Cevdet Kaya; Mustafa Güneş; Metin Öztürk; Muhammet İhsan Karaman
Journal:  Turk J Urol       Date:  2014-09

7.  [The 80 W potassium-titanium-phosphate (KTP) laser vaporization of the prostate. Technique and 6 month follow-up after 70 procedures].

Authors:  A Bachmann; O Reich; St Wyler; R Ruszat; R Casella; T Gasser; A Hofstetter; T Sulser
Journal:  Urologe A       Date:  2004-10       Impact factor: 0.639

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.