Literature DB >> 95280

Suppression of an external demand pacemaker by diaphragmatic myopotentials: a sign of electrode perforation?

M El Gamal, B Van Gelder.   

Abstract

Transient pacemaker inhibition was observed in a patient with an external bipolar demand pulse generator. Recordings of the proximal, distal and bipolar electrograms showed extracardiac myopotentials only in the unipolar distal and bipolar configuration. These myopotentials which caused inhibition of the pulse generator were related to respiration and other maneuvers that resulted in active contraction of the diaphragm, thus proving their diaphragmatic origin.

Entities:  

Mesh:

Year:  1979        PMID: 95280     DOI: 10.1111/j.1540-8159.1979.tb05200.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Inappropriate arrhythmia detection in implantable defibrillator therapy due to oversensing of diaphragmatic myopotentials.

Authors:  B Schulte; J Sperzel; J Carlsson; M Dürsch; A Erdogan; H F Pitschner; J Neuzner
Journal:  J Interv Card Electrophysiol       Date:  2001-12       Impact factor: 1.900

  1 in total