Mainstream end-tidal carbon dioxide monitoring in the neonatal intensive care unit.

H J Rozycki1, G D Sysyn, M K Marshall, R Malloy, T E Wiswell.   

Abstract

BACKGROUND: Continuous noninvasive monitoring of arterial carbon dioxide (CO2) in neonatal intensive care unit (NICU) patients would help clinicians avoid complications of hypocarbia and hypercarbia. End-tidal CO2 monitoring has not been used in this population to date, but recent technical advances and the introduction of surfactant therapy, which improves ventilation-perfusion matching, might improve the clinical utility of end-tidal monitoring.
OBJECTIVE: To determine the accuracy and precision of end-tidal CO2 monitoring in NICU patients.
DESIGN: Nonrandomized recording of simultaneous end-tidal and arterial CO2 pairs.
SETTING: Two university NICUs. PATIENTS: Forty-five newborn infants receiving mechanical ventilation who had indwelling arterial access, and a predefined subsample of infants who were <1000 g birth weight, <8 days of age, and who received surfactant therapy (extremely low birth weight -ELBW- <8). OUTCOME MEASURES: The correlation coefficient, degree of bias, and 95% confidence interval were determined for both the overall population and the ELBW <8 subgroup. Those factors which significantly influenced the bias were identified. The ability of the end-tidal monitor to alert the clinician to instances of hypocarbia or hypercarbia was determined.
RESULTS: There were 411 end-tidal/arterial pairs analyzed from 45 patients. The correlation coefficient was 0.833 and the bias was -6. 9 mm Hg (95% confidence interval, +/-11.5 mm Hg). The results did not differ markedly in the ELBW <8 infants. Measures of the degree of lung disease, the ventilation index and the oxygenation index, had small influences on the degree of bias. This type of capnometry identified 91% of the instances when the arterial CO2 pressure was between 34 and 54 mm Hg using an end-tidal range of 29 to 45 mm Hg. End-tidal values outside this range had a 63% accuracy in predicting hypocarbia or hypercarbia.
CONCLUSION: End-tidal CO2 monitoring in NICU patients is as accurate as capillary or transcutaneous monitoring but less precise than the latter. It may be useful for trending or for screening patients for abnormal arterial CO2 values.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9521950     DOI: 10.1542/peds.101.4.648

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  18 in total

1.  Continuous End-Tidal Carbon Dioxide Monitoring during Resuscitation of Asphyxiated Term Lambs.

Authors:  Praveen Kumar Chandrasekharan; Munmun Rawat; Jayasree Nair; Sylvia F Gugino; Carmon Koenigsknecht; Daniel D Swartz; Payam Vali; Bobby Mathew; Satyan Lakshminrusimha
Journal:  Neonatology       Date:  2016-02-12       Impact factor: 4.035

Review 2.  Are carbon dioxide detectors useful in neonates?

Authors:  E J Molloy; K Deakins
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2006-07       Impact factor: 5.747

Review 3.  [Anesthetic management of surgery in term and preterm infants].

Authors:  C Breschan; R Likar
Journal:  Anaesthesist       Date:  2006-10       Impact factor: 1.041

4.  Acceptable respiratory physiologic limits for children during weaning from mechanical ventilation.

Authors:  Miriam Santschi; France Gauvin; Georges Hatzakis; Jacques Lacroix; Philippe Jouvet
Journal:  Intensive Care Med       Date:  2006-10-25       Impact factor: 17.440

5.  Monitoring of end tidal carbon dioxide and transcutaneous carbon dioxide during neonatal transport.

Authors:  D G Tingay; M J Stewart; C J Morley
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2005-04-29       Impact factor: 5.747

Review 6.  Perioperative central nervous system injury in neonates.

Authors:  M E McCann; S G Soriano
Journal:  Br J Anaesth       Date:  2012-12       Impact factor: 9.166

7.  Detection of carbon dioxide thresholds using low-flow sidestream capnography in ventilated preterm infants.

Authors:  Emmanuel Lopez; Sophie Grabar; Alexandre Barbier; Baruch Krauss; Pierre-Henri Jarreau; Guy Moriette
Journal:  Intensive Care Med       Date:  2009-11       Impact factor: 17.440

8.  Minimizing blood loss and the need for transfusions in very premature infants.

Authors:  Brigitte Lemyre; Megan Sample; Thierry Lacaze-Masmonteil
Journal:  Paediatr Child Health       Date:  2015 Nov-Dec       Impact factor: 2.253

9.  Transcutaneous PCO2 monitoring in infants hospitalized with viral bronchiolitis.

Authors:  S Gal; A Riskin; I Chistyakov; N Shifman; I Srugo; A Kugelman
Journal:  Eur J Pediatr       Date:  2014-08-28       Impact factor: 3.183

10.  The correlation and level of agreement between end-tidal and blood gas pCO2 in children with respiratory distress: a retrospective analysis.

Authors:  James M Moses; Jamin L Alexander; Michael Sd Agus
Journal:  BMC Pediatr       Date:  2009-03-12       Impact factor: 2.125

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.