Continuous flow immunosensor for atrazine detection.

F Vianello1, L Signor, A Pizzariello, M L Di Paolo, M Scarpa, B Hock, T Giersch, A Rigo.   

Abstract

The hapten atrazine was detected under continuous flow conditions using a micro-column which contained immobilized monoclonal antibodies (Ab) against atrazine and atrazine labeled with alkaline phosphatase (An*). The equilibrium of the antibody-hapten system, was achieved by a continuous flow of the tracer An* through the micro-column containing the immobilized antibodies. The activity of the tracer was monitored continuously, after the micro-column, by an amperometric detector using p-hydroquinone phosphate as substrate. When pulses of unlabeled atrazine (An) were added to the An* flowing continuously through the micro-column, An* bound to the antibody was displaced, with a consequent change of the detector signal. By this method atrazine concentrations in the range 9-180 micrograms/l were monitored under conditions of continuous operation. Since the equilibrium condition for the system Ab-An* was continuously restored by the flow of An* through the micro-column the regeneration of the antibody was not required.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9519450     DOI: 10.1016/s0956-5663(97)00079-1

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  2 in total

Review 1.  Environmental toxicity monitoring using electrochemical biosensing systems.

Authors:  S Solé; S Alegret
Journal:  Environ Sci Pollut Res Int       Date:  2001       Impact factor: 4.223

2.  Single-walled carbon nanotube-based chemiresistive affinity biosensors for small molecules: ultrasensitive glucose detection.

Authors:  Lakshmi N Cella; Wilfred Chen; Nosang V Myung; Ashok Mulchandani
Journal:  J Am Chem Soc       Date:  2010-04-14       Impact factor: 15.419

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.