| Literature DB >> 9519447 |
S Sasaki1, I Karube, N Hirota, Y Arikawa, M Nishiyama, M Kukimoto, S Horinouchi, T Beppu.
Abstract
The enzymatic reaction of nitrite reductase (NIR) from Alcaligenes faecalis S-6 was applied to the measurement of nitrite. NIR was immobilized on the surface of a gold electrode using filter paper and a dialysis membrane, and used as a working electrode in a three-electrode system. Amperometric methods were applied using NIR and the electron mediator 1-methoxy PMS (1-methoxy-5-methylphenazinium methylsulfate). The decrease in cathodic current showed a correlation to nitrite concentration over the range 0-1 mg/l. Measurements using a batch-flow type system gave a lower detection limit of 0.01 mg/l. This is sufficient for the detection of nitrite in natural waters.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9519447 DOI: 10.1016/s0956-5663(97)00100-0
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618