The value of the coefficient of variation in assessing repeat variation in ECG measurements.

S C McLaughlin1, T C Aitchison, P W Macfarlane.   

Abstract

AIMS: The coefficient of variation is a popular measure for describing the amount of repeat variability present in ECG measurements from recording to recording. However, it can be misleading. The aim of the present study was to assess repeat variation (reclassification) in computer measured ECG criteria, i.e. positive to negative or vice versa, and compare this with the coefficient of variability. METHODS AND
RESULTS: Two ECGs were obtained from each of 295 patients, one day apart, and separately from a further 364 patients, several minutes apart. All patients were considered to be in a stable condition. Estimates of the coefficients of variation were obtained for a number of ECG parameters used in the diagnosis of left ventricular hypertrophy. Corresponding reclassification rates of relevant ECG criteria were also calculated. Large coefficients of variation were observed in voltage parameters, e.g. R in V5 (20% for day-to-day recordings and 6% for minute-to-minute recordings) while the corresponding reclassification rates were 8% and 0% respectively. The repeat variation in the diagnosis of left ventricular hypertrophy was up to 5% for day-to-day recordings and up to 3% for minute-to-minute recordings based on several different criteria.
CONCLUSION: A large coefficient of variation in a particular variable does not necessarily correspond to a high reclassification rate. A better measure of the impact of ECG variability for a particular measurement is obtained from its reclassification rate. In turn, this may have a minimal effect on the overall diagnosis of a particular abnormality.

Entities:  

Mesh:

Year:  1998        PMID: 9519330     DOI: 10.1053/euhj.1997.0741

Source DB:  PubMed          Journal:  Eur Heart J        ISSN: 0195-668X            Impact factor:   29.983


  3 in total

1.  Comprehensive brain analysis with automated high-resolution magnetization transfer measurements.

Authors:  Ying Wu; Hongyan Du; Pippa Storey; Christopher Glielmi; Fiona Malone; Shawn Sidharthan; Ann Ragin; Paul S Tofts; Robert R Edelman
Journal:  J Magn Reson Imaging       Date:  2011-10-11       Impact factor: 4.813

2.  Comparison of the Lunar Prodigy and iDXA Dual-Energy X-ray Absorptiometers for Assessing Total and Regional Body Composition.

Authors:  Shannon A Morrison; Robert M Petri; Heather L Hunter; Dheeraj Raju; Barbara Gower
Journal:  J Clin Densitom       Date:  2015-07-21       Impact factor: 2.617

3.  Short-term repeatability of electrocardiographic criteria of left ventricular hypertrophy.

Authors:  Michelle L Meyer; Elsayed Z Soliman; Dominique Drager; Gerardo Heiss
Journal:  Ann Noninvasive Electrocardiol       Date:  2019-09-09       Impact factor: 1.468

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.