Abstract

An automatic control system has been developed to balance and control the output of an artificial heart. The system consisted of 2 linked negative feedback loops. The left ventricle was controlled by a Servo-Stroke Optimizer, which insured complete filling and full stroke operation of the left pump with each beat and changed the beat rate in accord with changes in aortic pressure. The right ventricle was controlled by a Servo-Variable Systolic Duration unit, which changed the stroke of the right ventricle to maintain the left atrial pressure within a preset band. In the initial animal studies, control pressures (aortic and left atrial pressure) were obtained from implanted transducers. More recently, a method has been devised whereby the control pressures were obtained from specific points on the left air line pressure wave, obviating the need for implanted transducers. The control system has been evaluated in a mock circulatory loop and in 9 calves with implanted artificial hearts. The system has provided balance of the 2 implanted ventricles and changes in flow rate in response to changes in peripheral resistance. The need for an operator to make manual adjustment to the power units has been considerably reduced. Further studies are indicated to evaluate the function of the control system in animals performing moderate-to-maximal exercise. Furthermore, the ultimate benefits of full stroke operation vs fill limited mode operation remain to be delineated.

Entities:  

Mesh:

Year:  1976        PMID: 951850

Source DB:  PubMed          Journal:  Trans Am Soc Artif Intern Organs        ISSN: 0066-0078


  3 in total

1.  THE DESIGN AND FABRICATION OF A NEW TOTAL ARTIFICIAL HEART.

Authors:  Kevin Cheng; James W. Meador; Miguel A. Serrato; Tetsuzo Akutsu
Journal:  Cardiovasc Dis       Date:  1977

2.  Appropriate control time constant in relation to characteristics of the baroreflex vascular system in 1/R control of the total artificial heart.

Authors:  Sora Mizuta; Itsuro Saito; Takashi Isoyama; Shintaro Hara; Terumi Yurimoto; Xinyang Li; Haruka Murakami; Toshiya Ono; Kunihiko Mabuchi; Yusuke Abe
Journal:  J Artif Organs       Date:  2017-05-17       Impact factor: 1.731

3.  Concept of left atrial pressure estimation using its pulsatile amplitude in the helical flow total artificial heart.

Authors:  Sheng-Yuan Wu; Itsuro Saito; Takashi Isoyama; Yusuke Inoue; Masami Sato; Shintaro Hara; Xin-Yang Li; Terumi Yurimoto; Haruka Murakami; Yukino Kawase; Toshiya Ono; Yusuke Abe
Journal:  J Artif Organs       Date:  2014-09-03       Impact factor: 1.731

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.