| Literature DB >> 9517558 |
M E Legget1, G Bashein, J A McDonald, B I Munt, R W Martin, C M Otto, F H Sheehan.
Abstract
Ten phantoms were scanned with a multiplane transesophageal echocardiographic probe in a water bath to assess a new method for three-dimensional modeling of the mitral annulus. The annulus was reconstructed from manually outlined borders with Fourier series in each of the three spatial coordinates. Comparisons with direct measurements by least-squares linear regression gave coefficients of determination of 0.99 for annular height, area, and circumference. Expressed as a percentage of their true values, the mean +/- SD of the errors were -0.1% +/- 3.0% for annular height, -2.8% +/- 3.1% for area, and -0.2% +/- 1.7% for circumference. The mean residual error length for phantoms was 0.64 mm compared with 1.21 mm in nine patients studied during general anesthesia. This method gives accurate and precise measurements of the mitral annulus in vitro and should be valuable for studying its morphology and dynamics in vivo.Entities:
Mesh:
Year: 1998 PMID: 9517558 DOI: 10.1016/s0894-7317(98)70076-8
Source DB: PubMed Journal: J Am Soc Echocardiogr ISSN: 0894-7317 Impact factor: 5.251