J Swanpalmer1, R Kullenberg, T Hansson.
Abstract
Our laboratory has previously reported a method of determining the amount of bone mineral using triple-energy absorptiometry with a continuous x-ray spectrum. In the present study, the experimental properties of the technique were examined. The accuracy, the influence of fat content and body thickness and the in vitro and in vivo precision were analysed. The results found in this investigation showed that despite the complexity of the technique, the amount of bone mineral can be accurately determined. The in vivo precision was determined to be 3.4%, expressed as the coefficient of variation (CV), for different skeletal parts. The in vitro precision was found to be 2.1% (CV). Neither the fat content nor the body thickness had any effect on the measured bone mineral values. Excellent linearity and a close correlation were found between the true and the measured bone mineral values.Mesh:
Substances:
Year: 1998 PMID: 9509533 DOI: 10.1088/0031-9155/43/2/012
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609