Literature DB >> 9507555

Implantation of active can implantable defibrillators in the right pectoral region.

S M Jensen1, A Pietersen, X Chen.   

Abstract

Routinely the active can ICD is placed in the left side pectoral position, which theoretically gives optimal conditions for a low defibrillation threshold. Some patients, however, demand a right pectoral position, which possibly could result in a higher defibrillation threshold. A right pectoral position was used in 3 of 85 active can ICDs implanted in our institution from 1994. The DFT was 12 J in two and 18 J in one patient. Thus, right pectoral implantation is feasible and offers an alternative approach in selected patients.

Entities:  

Mesh:

Year:  1998        PMID: 9507555     DOI: 10.1111/j.1540-8159.1998.tb00078.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Endovascular extraction techniques: Part 3: Results and indications in patients with an ICD.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-06       Impact factor: 2.380

2.  Increased defibrillation threshold with right-sided active pectoral can.

Authors:  P R Roberts; S Allen; T Betts; J M Morgan; J F Urban; T Whitman; D E Euler; M J Kallok
Journal:  J Interv Card Electrophysiol       Date:  2000-04       Impact factor: 1.900

3.  Significance of intraoperative testing in right-sided implantable cardioverter-defibrillators.

Authors:  Andreas Keyser; Michael K Hilker; Ekrem Ucer; Sigrid Wittmann; Christof Schmid; Claudius Diez
Journal:  J Cardiothorac Surg       Date:  2013-04-11       Impact factor: 1.637

  3 in total