Abstract
Clinical observations have revealed that diode readings from 15 MV photon beams with small field sizes and small SSD values were consistently 5%-10% higher than doses calculated at the maximum depth. Dose at depth depends on the collimator scattering factor and the phantom scattering factor--both field size dependent. Diodes are placed on the skin surface, and therefore the phantom scatter factor is different than at depth. In addition, inverse square corrections should be calculated with respect to the location of the diode itself rather than the location of Dmax within the tissue. This may account for about 8% difference between calculated Dmax and diode readings for very small radiation fields and small SSDs, e.g., a prostate boost.Mesh:
Year: 1998 PMID: 9507487 DOI: 10.1118/1.598192
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071