Abstract
Percentage depth dose data acquired during the revision of BJR Supplement 17: Central Axis Depth Dose Data for Use in Radiotherapy have been reviewed. Coverage of the half value layer (HVL) and field diameter rang was broad, although gaps were apparent at HVLs of 3-5 and 6-7 mm A1 and field diameters > 14 cm. Comparisons of submitted data with that in Supplement 17 indicate a generally even spread of data about the existing PDD values. Most data agreed to within +/- 5% at 2 cm and +/- 15% at 10 cm depth. Systematic discrepancies may be due to the use of different phantom materials. There is evidence that the spread of data is at least partially due to the problems of beam quality specification. The inadequacy of the first HVL as a specifier of beam quality is a well known and a more realistic specifier might be defined in terms of a broad beam dose ratio at defined depths, for a given beam geometry.Mesh:
Year: 1997 PMID: 9498916 DOI: 10.1259/bjr.70.837.9498916