Iothalamate (IOT) clearance and p-aminohippuric acid (PAH) clearance are used for estimation of glomerular filtration rate (GFR) and effective renal plasma flow (ERPF). A simple and rapid method is described for simultaneous determination of IOT and PAH in the same chromatogram in the serum and urine of humans. The mobile phase consisted of methanol-50 mM sodium monobasic phosphate with 0.5 mM tetrabutyl ammonium hydrogen sulfate (18:82, v/v), pumped at a rate of 0.8 ml/min on a C18 reversed-phase column. Samples of serum and urine were deproteinized with two volumes of acetonitrile containing the internal standard, p-aminobenzoic acid (PABA). The UV detector was set at 254 nm and peak height ratios of PAH or IOT to PABA were calculated with an integrator. Precision and accuracy were within 15% for both PAH and IOT. The recovery of PAH in urine and serum were 94% and 91%, respectively. For IOT the corresponding recoveries were 93% and 92%, respectively. This method clearly distinguishes acetyl-PAH from PAH and has been validated in healthy volunteers.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9498664 DOI: 10.1016/s0378-4347(97)00497-0
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273