Oxygen microelectrode response to oxygen pressures up to 15 ATA.

J S Colton, J A Michael, C A Colton.   

Abstract

Polarographic measurements of oxygen at high pressures were made at 21-22 degrees C with open-type Wood's metal microelectrodes with tip diameter about 5 mu. Although the sensitivity of the individual electrodes ranged from 3.1 x 10(-10) to 8.8 x 10(-12) amps/mm Hg of O2, the oxygen current was always a straight line function of Po2 from 21% O2 at ambient pressure (Po2 = 160 mm Hg) to 100% O2 at 200 psig (Po2 = 11,102 mm Hg). Compression to 200 psig with 100% nitrogen produced no significant change in microelectrode current, showing that the microelectrodes are insensitive to pressure alone.

Entities:  

Mesh:

Substances:

Year:  1976        PMID: 949302