Reproducibility of brachial ultrasonography and flow-mediated dilatation (FMD) for assessing endothelial function.

K L Hardie1, S Kinlay, D B Hardy, J Wlodarczyk, J S Silberberg, P J Fletcher.   

Abstract

BACKGROUND: High-resolution brachial artery ultrasonography is used to study vasodilator response induced by physiologic reactive hyperaemia. We examined the reproducibility of measuring flow-mediated dilatation (FMD) on two occasions. AIMS: To determine the degree of variability of this technique in our vascular laboratory for the design of clinical research studies.
METHODS: Nineteen subjects were studied on two separate occasions using an Acuson 128 ultrasound device and a 7.0 MHz linear array transducer. Reactive hyperaemia was induced in the brachial artery by inflation and release of a blood pressure cuff. Nitrate-induced dilatation was assessed in 11 of the 19 subjects. Measurements were made by two observers blinded to subject details.
RESULTS: The 11 subjects given sublingual GTN during the first ultrasound study had a mean nitrate-induced dilatation of 20.7% (sd 9.6). The mean vessel diameter of 3.78 mm (sd 0.7) at rest and 3.89 mm (sd 0.7) during reactive hyperaemia yielded a mean FMD of only 3.0% (sd 2.7). The mean difference in FMD within-observers was 0.13% (sd 2.07), between-observers 0.06% (sd 2.17) and between-studies was 0.57% (sd 6.83).
CONCLUSIONS: The reproducibility of FMD measured by brachial artery ultrasound was poor and likely to render the measurements inaccurate for clinical research in our hands. Between-study variation contributed the largest proportion of total study variability. We suggest that investigators using this technique conduct their own careful reproducibility studies in order to avoid the misinterpretation of 'negative' studies.

Entities:  

Mesh:

Year:  1997        PMID: 9483230     DOI: 10.1111/j.1445-5994.1997.tb00992.x

Source DB:  PubMed          Journal:  Aust N Z J Med        ISSN: 0004-8291


  6 in total

Review 1.  Venous occlusion plethysmography in cardiovascular research: methodology and clinical applications.

Authors:  I B Wilkinson; D J Webb
Journal:  Br J Clin Pharmacol       Date:  2001-12       Impact factor: 4.335

2.  Reliability of ultrasonography in detecting flow mediated dilation.

Authors:  Selcuk Pala; Gokhan Kahveci; Mustafa Akcakoyun; Ramazan Kargin; Kursat Tigen; Ali Cevat Tanalp; Tansu Karaahmet; Hekim Karapinar; Cevat Kirma
Journal:  Int J Cardiovasc Imaging       Date:  2009-07-17       Impact factor: 2.357

Review 3.  Flow-mediated dilatation.

Authors:  O T Raitakari; D S Celermajer
Journal:  Br J Clin Pharmacol       Date:  2000-11       Impact factor: 4.335

4.  Similar pattern of change in V̇o2 kinetics, vascular function, and tissue oxygen provision following an endurance training stimulus in older and young adults.

Authors:  Kaitlin M McLay; Juan M Murias; Donald H Paterson
Journal:  Am J Physiol Regul Integr Comp Physiol       Date:  2017-01-25       Impact factor: 3.619

Review 5.  New Insights From MRI and Cell Biology Into the Acute Vascular-Metabolic Implications of Electronic Cigarette Vaping.

Authors:  Felix W Wehrli; Alessandra Caporale; Michael C Langham; Shampa Chatterjee
Journal:  Front Physiol       Date:  2020-05-21       Impact factor: 4.566

6.  Repeatability of vascular responsiveness measures derived from near-infrared spectroscopy.

Authors:  Kaitlin M McLay; Josh P Nederveen; Silvia Pogliaghi; Donald H Paterson; Juan M Murias
Journal:  Physiol Rep       Date:  2016-05
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.