Literature DB >> 9477785

Periotest values of dental implants in the first 2 years after second-stage surgery: DICRG interim report no. 8. Dental Implant Clinical Research Group.

L Walker1, H F Morris, S Ochi.   

Abstract

In 1991, the Dental Implant Clinical Research Group initiated a long-term clinical study in cooperation with the Department of Veterans Affairs to investigate the influence of implant design, application, and site of placement on clinical performance and crestal bone height. As part of this investigation, Periotest values for 2,212 root from implants were determined at second-stage surgery and during a 24-month follow-up period. Mean Periotest values decreased for implants placed in quality 1 and 2 bone, did not change for implants in quality 3 bone, and increased for implants in quality 4 bone. Implants in the posterior maxilla and single implants in the anterior maxilla had increasing mean Periotest values as compared with decreasing values for implants in other regions. Mean Periotest values for uncoated implants decreased gradually to approach those of hydroxyapatite-coated implants.

Entities:  

Mesh:

Year:  1997        PMID: 9477785     DOI: 10.1097/00008505-199700630-00007

Source DB:  PubMed          Journal:  Implant Dent        ISSN: 1056-6163            Impact factor:   2.454


  2 in total

1.  Assessment of Osstell™ and Periotest® systems in measuring dental implant stability (in vitro study).

Authors:  Samer Al-Jetaily; Abdullah Alfarraj Al-Dosari
Journal:  Saudi Dent J       Date:  2010-11-13

2.  Implant Stability Changes for Nonsubmerged and Submerged Protocols for a Single Implant Mandibular Overdenture Using Ball Attachment.

Authors:  Ahmed Salah; Karim Foda; Mohamed Farouk Abdalla; Marwa Abdel Aal; Amr Naguib; Nouran Abdel Nabi
Journal:  Int J Dent       Date:  2021-09-16
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.