| Literature DB >> 9474801 |
K A Barnes1, R A Smith, K Williams, A P Damant, M J Shepherd.
Abstract
A microbore high performance liquid chromatographic/electrospray/mass spectrometric (HPLC/ESI-MS) method has been developed for the determination of the phytoestrogens daidzein and genistein in soya flours and baby foods. The samples were hydrolysed and extracted with acetonitrile-water prior to analysis. LC was performed on a microbore Primesphere 5C8 column using a water/acetonitrile/acetic acid mobile phase at a flow rate of 60 microliters/min. Atmospheric pressure ionization in the form of pneumatically assisted electrospray mass spectrometry (ESI-MS) was used as the method of detection. The limit of detection was 0.2 mg/ kg for daidzein and 0.7 mg/kg for genistein in the flour and food samples. The method proved both robust and reliable when operated over a long time period (10 days, 463 injections) generating precision data with a coefficient of variation of 4-15%.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9474801 DOI: 10.1002/(SICI)1097-0231(19980214)12:3<130::AID-RCM127>3.0.CO;2-9
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419