Literature DB >> 9474434

Limitations of the MMS-10 tympanic displacement analyser.

H J Rosingh1, H P Wit, F W Albers.   

Abstract

Using the MMS-10 tympanic displacement analyser, the test-retest variation was determined in 20 normal hearing ears (ten subjects) for the measurement parameters Vi and Vm. Variations were found to be on average 60 nl for Vi and 72 nl for Vm. The consequences of these substantial test-retest variabilities for the applicability of the device and interpretation of the measurement results are discussed.

Mesh:

Year:  1998        PMID: 9474434     DOI: 10.3109/00206099809072956

Source DB:  PubMed          Journal:  Audiology        ISSN: 0020-6091


  2 in total

1.  Distortion product otoacoustic emissions for assessment of intracranial hypertension at extreme altitude?

Authors:  Bernhard Olzowy; Gregor von Gleichenstein; Martin Canis; Klaus Mees
Journal:  Eur J Appl Physiol       Date:  2008-01-11       Impact factor: 3.078

2.  Posture systematically alters ear-canal reflectance and DPOAE properties.

Authors:  Susan E Voss; Modupe F Adegoke; Nicholas J Horton; Kevin N Sheth; Jonathan Rosand; Christopher A Shera
Journal:  Hear Res       Date:  2010-03-19       Impact factor: 3.208

  2 in total