Three-dimensional electrophysiological imaging of the intact canine left ventricle using a noncontact multielectrode cavitary probe: study of sinus, paced, and spontaneous premature beats.

D S Khoury1, K L Berrier, S M Badruddin, W A Zoghbi.   

Abstract

BACKGROUND: The feasibility of measuring cavitary electrograms using a noncontact probe and reconstructing endocardial surface electrograms and activation sequences during paced beats was previously demonstrated in the isolated canine left ventricle (LV). The objective of the present study was to develop and test a high-resolution, three-dimensional, endocardial electrophysiological imaging technique that simultaneously reconstructs endocardial surface electrograms and their corresponding activation sequences during normal and abnormal beats with the use of cavitary electrograms measured with a noncontact multielectrode probe in the intact canine LV. METHODS AND
RESULTS: A 128-electrode probe was inserted into the intact canine LV. Probe unipolar electrograms were simultaneously acquired during sinus, artificially paced, and spontaneous premature beats. Representative endocardial electrograms were measured directly using eight needle electrodes (the "gold standard"). A probe-cavity realistic, three-dimensional geometric model was constructed using two-dimensional epicardial echocardiography. Boundary element methods and numeric regularization were used to compute electrograms at 194 sites on the endocardium. In eight pacing protocols, computed endocardial electrograms correlated well with directly measured electrograms (r=.88). Corresponding activation times were also in agreement with those determined from measured endocardial electrograms (activation error, 4.7 ms). The earliest region of activation was invariably in the vicinity of the pacing needle (spatial error, 9.2 mm). Subsequently, the site of origin of ischemia-induced spontaneous ventricular premature beats and the ensuing sequence of depolarization was identified.
CONCLUSIONS: Noncontact mapping provides realistic, three-dimensional electrophysiological images of the endocardium, on a beat-by-beat basis, that localize the sites of origin of premature beats and reconstruct their activation sequences.

Entities:  

Mesh:

Year:  1998        PMID: 9468214     DOI: 10.1161/01.cir.97.4.399

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  11 in total

Review 1.  Recent advances in cardiac mapping techniques.

Authors:  C Schmitt; G Ndrepepa; I Deisenhofer; M Schneider
Journal:  Curr Cardiol Rep       Date:  1999-07       Impact factor: 2.931

2.  Stereotactic magnetic resonance guidance for anatomically targeted ablations of the fossa ovalis and the left atrium.

Authors:  Timm Dickfeld; Hugh Calkins; Menekhem Zviman; Glenn Meininger; Lars Lickfett; Ariel Roguin; Albert C Lardo; Ronald Berger; Henry Halperin; Stephen B Solomon
Journal:  J Interv Card Electrophysiol       Date:  2004-10       Impact factor: 1.900

3.  Application of the method of fundamental solutions to potential-based inverse electrocardiography.

Authors:  Yong Wang; Yoram Rudy
Journal:  Ann Biomed Eng       Date:  2006-06-29       Impact factor: 3.934

Review 4.  Magnetic resonance imaging and radiofrequency ablations.

Authors:  T Dickfeld
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2007-09

5.  Can noncontact mapping distinguish between endo- and epicardial foci?

Authors:  Frederik Voss; Alexander Bauer; Steffen Witte; Hugo A Katus; Ruediger Becker
Journal:  Clin Res Cardiol       Date:  2008-05-03       Impact factor: 5.460

6.  Transseptal activation during left atrial pacing in humans: electroanatomic mapping using a noncontact catheter and the intracardiac echocardiography.

Authors:  Leonardo Calò; Filippo Lamberti; Maria Luisa Loricchio; Antonio Castro; Augusto Boggi; Furio Colivicchi; Claudio Pandozi; Massimo Santini
Journal:  J Interv Card Electrophysiol       Date:  2002-06       Impact factor: 1.900

Review 7.  Catheter ablation of ventricular tachycardia related to coronary artery disease: the role of noncontact mapping.

Authors:  A W Chow; R J Schilling; N S Peters; D W Davies
Journal:  Curr Cardiol Rep       Date:  2000-11       Impact factor: 2.931

8.  Imaging cardiac activation sequence during ventricular tachycardia in a canine model of nonischemic heart failure.

Authors:  Chengzong Han; Steven M Pogwizd; Long Yu; Zhaoye Zhou; Cheryl R Killingsworth; Bin He
Journal:  Am J Physiol Heart Circ Physiol       Date:  2014-11-21       Impact factor: 4.733

Review 9.  Non-contact mapping in the treatment of ventricular tachycardia after myocardial infarction.

Authors:  Kim Rajappan; Richard J Schilling
Journal:  J Interv Card Electrophysiol       Date:  2007-06-28       Impact factor: 1.900

10.  Real-time cardiac catheter navigation on three-dimensional CT images.

Authors:  Stephen B Solomon; Timm Dickfeld; Hugh Calkins
Journal:  J Interv Card Electrophysiol       Date:  2003-02       Impact factor: 1.900