| Literature DB >> 9466787 |
Y Carmeli1, K Eichelberger, D Soja, J Dakos, L Venkataraman, P DeGirolami, M Samore.
Abstract
False results showing an outbreak of Pseudomonas aeruginosa with resistance to imipenem were traced to a defective lot of microdilution MIC testing panels. These panels contained two- to threefold lower concentrations of imipenem than expected and resulted in artifactual two- to fourfold increases in MICs of imipenem. The quality-control MIC results for Pseudomonas aeruginosa ATCC 27853 were 4 microg/ml, the highest value within the range recommended by the National Committee for Clinical Laboratory Standards. We recommend that this value be considered out of the quality-control range.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9466787 PMCID: PMC104588
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948