A Martin-Esteban1, P Fernández, D Stevenson, C Cámara.
Abstract
An immunosorbent containing antiisoproturon and antichlortoluron antibodies immobilised on aldehyde-activated silica was employed for the on-line preconcentration and liquid chromatography-diode array detection of several phenylureas. The efficiency of the coupling of the immunosorbent to the liquid chromatographic system and the characteristics of the immunosorbent were evaluated. The on-line system allowed the selective trace enrichment of chlortoluron, isoproturon, metobromuron, linuron and chlorbromuron at the 0.05-0.5 microgram l-1 level in ground and river waters and provided detection limits in the range 0.01-0.03 microgram l-1 by percolating only 10 ml of water sample. The proposed method was validated by analysing freeze-dried tap water samples with a high content in pesticides of different chemical functionalities. The results obtained were compared with the mean value obtained in an interlaboratory exercise.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9463965 DOI: 10.1039/a702828h