Wound infection rate and irrigation pressure of two potential new wound irrigation devices: the port and the cap.

J W Morse1, T Babson, C Camasso, A C Bush, P A Blythe.   

Abstract

The objectives of this study were to determine the speed of irrigation and the infection rate of two new irrigation devices. In the clinical portion of this two-part study, 208 patients with traumatic wounds were randomized to one of two new irrigation device groups, the cap/bottle or the port/bag, to determine irrigation times and infection/complication rates. Wounds were irrigated in less than 4 minutes in 97% of patients using a mean of 786 mL. The combined infection/complication rates were: port group, 1 of 99 (1%); cap group, 4 of 108 (4%), P = .356. In part II, 9 male and 8 female medical volunteers were timed in the delivery of 250 mL of saline into a graduated cylinder, using four different irrigation set-ups. Mean time for 250 mL and calculated stream pressures (psi) were: (1) cap/bottle, 12.9 seconds, 1.5 psi; (2) port/bag, 11.2 seconds, 2.0 psi; (3) syringe/catheter, 113.2 seconds, 8.2 psi; and (4) syringe/needle, 175.4 seconds, 7.3 psi (time and psi: P < .05 for all pairwise comparisons except cap versus port). Rapid irrigation and infection rates comparable with standard devices used in wound irrigation suggest that the new devices may prove to be valuable tools in emergent wound care.

Entities:  

Mesh:

Year:  1998        PMID: 9451311     DOI: 10.1016/s0735-6757(98)90062-6

Source DB:  PubMed          Journal:  Am J Emerg Med        ISSN: 0735-6757            Impact factor:   2.469


  2 in total

Review 1.  A critical review of irrigation techniques in acute wounds.

Authors:  Justin S Chatterjee
Journal:  Int Wound J       Date:  2005-09       Impact factor: 3.315

Review 2.  Intracavity lavage and wound irrigation for prevention of surgical site infection.

Authors:  Gill Norman; Ross A Atkinson; Tanya A Smith; Ceri Rowlands; Amber D Rithalia; Emma J Crosbie; Jo C Dumville
Journal:  Cochrane Database Syst Rev       Date:  2017-10-30
  2 in total