Large laser sheaths for pacing and defibrillator lead removal.

C Reiser1, K D Taylor, R A Lippincott.   

Abstract

BACKGROUND AND
OBJECTIVE: In a recent clinical trial, the 12-F laser sheath showed 95% success in completely explanting chronically implanted pacing leads smaller than 7.5-F diameter. For larger leads, two new sizes of laser sheath have been implemented, the 14-F and 16-F (outer diameter) devices, which accommodate leads up to 9.5- and 11.5-F, respectively. The object of this study was to determine the cutting ability of the larger devices compared to the 12-F design.
MATERIALS AND METHODS: The rate of device advancement through fresh porcine aorta was measured for three sizes of laser sheath as pulsed ultraviolet light from a 308-nm XeCl excimer laser was applied. Dependent variables were fluence (mJ/mm2), laser pulse repetition rate, and pressure applied between the device and the tissue.
RESULTS: At 60 mJ/mm2, 40 Hz repetition rate and 1.4 kg/cm2 pressure, all devices produced cutting rates in the range of 9-13 microns/pulse. Improvement in advancement per laser shot can be attained by increasing any independent variable studied.
CONCLUSIONS: Physicians must apply only slightly greater force to the larger laser sheaths, and maximum available repetition rate and fluence implies maximum cutting speed.

Entities:  

Mesh:

Year:  1998        PMID: 9443149     DOI: 10.1002/(sici)1096-9101(1998)22:1<42::aid-lsm10>3.0.co;2-a

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  5 in total

Review 1.  Advances in device lead extraction.

Authors:  C L Byrd
Journal:  Curr Cardiol Rep       Date:  2001-07       Impact factor: 2.931

Review 2.  Endovascular extraction techniques for pacemaker and ICD lead extraction: Part 1.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-04       Impact factor: 2.380

3.  A novel indication for transvenous lead extraction: upgrading implantable cardioverter defibrillator systems.

Authors:  V J Korley; N Hallet; M Daoust; L M Epstein
Journal:  J Interv Card Electrophysiol       Date:  2000-10       Impact factor: 1.900

4.  The lead extractor's toolbox: a review of current endovascular pacemaker and ICD lead extraction techniques.

Authors:  F A Bracke
Journal:  Indian Pacing Electrophysiol J       Date:  2003-07-01

5.  Comparison between laser sheaths, femoral approach and rotating mechanical sheaths for lead extraction.

Authors:  F A Bracke; N Rademakers; N Verberkmoes; M Van 't Veer; B M van Gelder
Journal:  Neth Heart J       Date:  2021-12-21       Impact factor: 2.854

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.