| Literature DB >> 9439459 |
F Frantzen1, K Grimsrud, D E Heggli, A L Faaren, T Løvli, E Sundrehagen.
Abstract
We present a new filter assay for the determination of glycohemoglobin as a unique application of the boronic acid affinity principle. With the use of a water-soluble blue-colored boronic acid derivative and a specific precipitation method for hemoglobin, total hemoglobin including bound boronic acid is precipitated and collected on a filter strip before quantification. Hemoglobin and boronic acid are quantified by a dual-wave-length reflectometric measurement, and the result is reported directly as percent glycohemoglobin. The test is simple, quick, and designed as a doctors' office test for the monitoring and management of diabetes. The imprecision of the assay is < 4% over the range 3-18% Hb A1c, and the method is linear up to at least 20% Hb A1c. Comparisons with four well-established glycohemoglobin methods yielded correlation coefficients ranging from 0.94 to 0.99, with slopes from 0.94 to 1.01.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9439459
Source DB: PubMed Journal: Clin Chem ISSN: 0009-9147 Impact factor: 8.327