Literature DB >> 9436559

Cardiac pacemaker infection: surgical management with and without extracorporeal circulation.

M J Wilhelm1, C Schmid, D Hammel, S Kerber, H M Loick, M Herrmann, H H Scheld.   

Abstract

BACKGROUND: Pacemaker infections are rare, but serious complications of pacemaker therapy. The generator pocket, the pacing leads, or both may be involved.
METHODS: We report on 12 patients with infected pacemaker systems. Four patients suffered from localized generator pocket infections, 6 had infected leads, and 2 patients had both. Pacemaker systems were completely removed in all patients. When the infection was limited to the generator pocket, the pacemaker system was removed at the original implantation site. Extracorporeal circulation was employed for the explantation of infected pacing leads.
RESULTS: No complications occurred in patients with localized generator pocket infections. One patient with infected leads who was preoperatively already in a serious clinical condition died of septic shock in the early postoperative period; another patient died of pulmonary complications after tricuspid valve replacement 14 months after pacemaker explantation. No recurrent infections were observed.
CONCLUSIONS: Explantation of the complete pacemaker system has proved a reliable method to eradicate infection. Complications have been rare, except in patients in a critically ill state who undergo cardiopulmonary bypass.

Entities:  

Mesh:

Year:  1997        PMID: 9436559     DOI: 10.1016/s0003-4975(97)00989-2

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  14 in total

1.  Successful removal of an infected pacemaker lead using cardiopulmonary bypass in an 89-year-old patient.

Authors:  Hiroomi Murayama; Takashi Watanabe; Naoki Kida; Takashi Yano; Keiji Ohara; Atsukata Kobayashi
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

Review 2.  [ICD lead defects: diagnosis and therapeutical options].

Authors:  Christian G Wollmann; Dirk Böcker; Andreas Löher; Hans H Scheld; Günter Breithardt; Rainer Gradaus
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-02-11

Review 3.  Endovascular extraction techniques for pacemaker and ICD lead extraction: Part 1.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-04       Impact factor: 2.380

Review 4.  Endovascular extraction techniques: Part 2: Complications and indications.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-05       Impact factor: 2.380

Review 5.  Surgical management of cardiac implantable electronic device infections.

Authors:  Michael Koutentakis; Stavros Siminelakis; Panagiotis Korantzopoulos; Anastasios Petrou; Alexandra Petrou; Helen Priavali; Eleftheria Priavali; Andreas Mpakas; Helen Gesouli; Eleftheria Gesouli; Efstratios Apostolakis; Eleftheria Apostolakis; Kosmas Tsakiridis; Paul Zarogoulidis; Nikolaos Katsikogiannis; Ioanna Kougioumtzi; Nikolaos Machairiotis; Theodora Tsiouda; Konstantinos Zarogoulidis
Journal:  J Thorac Dis       Date:  2014-03       Impact factor: 2.895

Review 6.  [Implant infections. What must an internist know?].

Authors:  W Zimmerli
Journal:  Internist (Berl)       Date:  2005-06       Impact factor: 0.743

7.  Successfully treated infected left ventricular pseudoaneurysm after the Dor operation.

Authors:  Takahiko Aoyama; Osamu Kawaguchi; Takao Teramoto; Chihiro Narumiya; Jin Hachiya; Yoshihisa Nagata
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2006-03

Review 8.  Candida infections of medical devices.

Authors:  Erna M Kojic; Rabih O Darouiche
Journal:  Clin Microbiol Rev       Date:  2004-04       Impact factor: 26.132

Review 9.  Micafungin for Candida albicans pacemaker-associated endocarditis: a case report and review of the literature.

Authors:  Carlo Tascini; Maria Grazia Bongiorni; Enrico Tagliaferri; Antonello Di Paolo; Sarah Flammini; Ezio Soldati; Alessandro Leonildi; Andrea Di Cori; Francesco Menichetti
Journal:  Mycopathologia       Date:  2012-10-17       Impact factor: 2.574

10.  Long-term outcome of complete cardiovascular implantable electronic device removal with cardiopulmonary bypass.

Authors:  Masaho Okada; Yuji Narita; Yoshimori Araki; Hideki Oshima; Akihiko Usui; Yuichi Ueda
Journal:  J Artif Organs       Date:  2013-01-20       Impact factor: 1.731

View more