O Niwa1, R Kurita, T Horiuchi, K Torimitsu.
Abstract
We report the first on-line electrochemical sensor for the continuous measurement of gamma-aminobutyric acid (GABA), which is a well-known inhibitory neurotransmitter in the nervous system. The sensor is composed of a glutamate oxidase (GluOx) and catalase immobilized small-volume enzymatic reactor and a glassy carbon (GC) electrode modified with a top layer film consisting of gabase and GluOx coimmobilized bovine serum albumin and an Ospoly(vinylpyrridine) bottom layer film containing horseradish peroxidase. The response of the sensor depends on the alpha-ketoglutarate concentration and is almost saturated when its concentration is 100 times higher than GABA. The sensor exhibits a sensitivity of 1.56 nA/microM for GABA under optimized conditions and shows almost no response when 10 microM glutamate is continuously injected. A detection limit of 0.1 microM is obtained with a linear range of 0.1-10 microM. GABA can be measured in the absence of alpha-ketoglutarate when there is L-glutamate in the sample solution, which is a typical condition for the extracellular measurement of cultured nerve cells.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9435468 DOI: 10.1021/ac970740z