Literature DB >> 9426413

Implanted programmable insulin pumps: one hundred fifty-three patient years of surgical experience.

R Udelsman1, H Chen, K Loman, H A Pitt, C D Saudek.   

Abstract

BACKGROUND: Implanted insulin pumps (IIPs) are an alternative treatment for diabetes mellitus. To maintain good glycemic control, patients with an IIP require frequent surgical interventions.
METHODS: Since November 1986, 21 patients with insulin-dependent diabetes mellitus at the Johns Hopkins Hospital have undergone implantation of pumps subcutaneously with a catheter delivering insulin into the peritoneal space. Patients were followed up with self-monitoring of blood glucose levels two to four times daily and percutaneous refills of the pump with U-400 insulin every 6 to 12 weeks.
RESULTS: In 153 patient-years, 77 pumps were placed. The mean pump life was 29 +/- 2 months. Morbidity was limited to pump site infections (3.9%) and one small bowel obstruction. Inadvertent insulin overdelivery has never occurred. Episodes of insulin underdelivery were caused by backflow anomalies (n = 67), which were cleared by percutaneous rinses, or catheter obstructions (n = 12), which required catheter replacements. Ninety-one operations were required to maintain pump function. All patients are alive and report improvements in quality of life. Significant reductions in hemoglobin A1C and plasma glucose levels were also seen.
CONCLUSIONS: Long-term use of IIPs results in significant improvements in clinical parameters and quality of life for individuals with insulin-dependent diabetes mellitus. Relatively frequent operations are required for maintaining pump function, which are done with a local anesthetic with minimal morbidity.

Entities:  

Mesh:

Year:  1997        PMID: 9426413     DOI: 10.1016/s0039-6060(97)90202-6

Source DB:  PubMed          Journal:  Surgery        ISSN: 0039-6060            Impact factor:   3.982


  3 in total

1.  A review of the management of implanted medical devices for diabetes: trends and directions.

Authors:  Carl Edman; Darrel Drinan
Journal:  J Diabetes Sci Technol       Date:  2008-11

2.  Surgical aspects and complications of continuous intraperitoneal insulin infusion with an implantable pump.

Authors:  Jan Willem Haveman; Susan J J Logtenberg; Nanne Kleefstra; Klaas H Groenier; Henk J G Bilo; Adri M Blomme
Journal:  Langenbecks Arch Surg       Date:  2008-12-02       Impact factor: 3.445

3.  Complications of continuous intraperitoneal insulin infusion with an implantable pump.

Authors:  Peter R van Dijk; Susan Jj Logtenberg; Klaas H Groenier; Jan Willem Haveman; Nanno Kleefstra; Henk Jg Bilo
Journal:  World J Diabetes       Date:  2012-08-15
  3 in total