Quality control aspects of the Philips multileaf collimator.

A R Hounsell1, T J Jordan.   

Abstract

BACKGROUND AND
PURPOSE: Linear accelerators equipped with multileaf collimators (MLCs) are becoming more common and are widely available from most commercial manufacturers. There is a need to ensure they retain their commissioning specification using a preventative maintenance and quality control (QC) programme. This paper considers the design criteria of the Philips MLC which are important to the production of a comprehensive quality control programme.
MATERIALS AND METHODS: The specific QC problems related to MLCs are identified as the positional accuracy of the leaves and their relationship to the back-up collimators, leakage considerations, the relationship of X-ray to light field and the influence of gravity on the positioning and leakage characteristics of the leaves. These problems are considered in relation to the general design considerations of the MLC, and methods of performing routine quality control checks are discussed. RESULTS AND
CONCLUSIONS: The introduction of MLCs into clinical use results in new QC procedures being developed but it can be concluded that for the Philips MLC only an extra 30 min of QC time is needed per month and that its use has added little to the general down-time of this department.

Entities:  

Mesh:

Year:  1997        PMID: 9426116     DOI: 10.1016/s0167-8140(97)00100-x

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  2 in total

1.  Evaluation of AutoCAL for electronic portal imaging device-based multi-leaf collimator quality assurance.

Authors:  Tarafder J Shameem
Journal:  Radiol Phys Technol       Date:  2015-10-27

2.  An Investigation of Multileaf Collimator Performance Dependence on Gantry Angle Using Machine Log Files.

Authors:  Jeremy L Hughes; Martin A Ebert; Conor K McGarry; Christina E Agnew; Mahsheed Sabet; Pejman Rowshanfarzad
Journal:  J Med Phys       Date:  2021-12-31
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.