Instrument performance in bone density testing at five Australian centres.

K M Khan1, S L Henzell, C Broderick, R L Prince, A Saul, J Lomman, J D Wark.   

Abstract

AIMS: To assess the in vitro precision and accuracy of bone mineral densitometry (BMD) within and between locations at five Australian centres.
METHODS: Using a multicentre reliability study the accuracy and short- and long-term precision of dual-energy X-ray absorptiometry (DXA) in vitro was compared on five instruments. Measures were performed using pencil beam mode on four Hologic QDR-2000 densitometers and one Hologic QDR-1000/W (Hologic Inc, Waltham, MA).
RESULTS: Short-term precision of bone mineral density measurement was less than 0.5% for spine phantoms (n = 10 for each centre, mean intrasite coefficient of variation [CV] 0.39 +/- 0.09% [SD]) and for hip phantoms (n = 10 for each centre, mean intrasite coefficient of variation [CV] 0.34 +/- 0.10% [SD]). Between-centre measurement (n = 10 for each phantom) of a single spine phantom and a single hip phantom (specified mineral contents-58.5 g and 38.6 g, respectively) revealed ranges of bone mineral content of 57.7-58.1 g (all-point CV = 0.52%) and 37.1-37.8 g (all-point CV = 0.70%), respectively. When results from pairs of machines were compared there were statistically different mean BMD results for the majority of the ten possible pairings for both spine and hip measurements. Each study centre measured in vitro stability of phantom BMD measurements over a one year period (n = 45-283, median 157 for spine; and n = 0-262, median 38, for hip); CVs ranged from 0.38 to 0.53% for the spine measurements and from 0.38 to 0.54% for the hip measurements. The mean all-point accuracy of the spine phantom measurements was 99.1% and the hip phantom measurements was 96.7%.
CONCLUSIONS: Across a number of instruments DXA demonstrates in vitro all-point precision of 0.5% for the spine phantom and 0.7% for the hip phantom. The instrument demonstrates accuracy of greater than 99% at the spine and 96% at the hip. This finding has clinical, research and quality control implications.

Entities:  

Mesh:

Year:  1997        PMID: 9404582     DOI: 10.1111/j.1445-5994.1997.tb00959.x

Source DB:  PubMed          Journal:  Aust N Z J Med        ISSN: 0004-8291


  4 in total

1.  A study of the long-term precision of dual-energy X-ray absorptiometry bone densitometers and implications for the validity of the least-significant-change calculation.

Authors:  T N Hangartner
Journal:  Osteoporos Int       Date:  2006-11-30       Impact factor: 4.507

2.  Self-reported ballet classes undertaken at age 10-12 years and hip bone mineral density in later life.

Authors:  K M Khan; K L Bennell; J L Hopper; L Flicker; C A Nowson; A J Sherwin; K J Crichton; P R Harcourt; J D Wark
Journal:  Osteoporos Int       Date:  1998       Impact factor: 5.071

3.  Whole-body dual-energy X-ray absorptiometry demonstrates better reliability than segmental body composition analysis in college-aged students.

Authors:  Petr Kutáč; Václav Bunc; Martin Sigmund
Journal:  PLoS One       Date:  2019-04-22       Impact factor: 3.240

4.  Effect of multiple injections of botulinum toxin into painful masticatory muscles on bone density in the temporomandibular complex.

Authors:  Karen G Raphael; Malvin N Janal; Aditya Tadinada; Vivian Santiago; David A Sirois; Alan G Lurie
Journal:  J Oral Rehabil       Date:  2020-09-19       Impact factor: 3.837

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.